HU-to-density calibration curves using diagnostic CT for radiotherapy treatment planning: a feasibility study
<i>Objective.</i>Accurate radiotherapy (RT) dose calculation critically depends on reliable conversion of Hounsfield units (HUs) to density, traditionally achieved through scanner-specific calibration using dedicated simulation computed tomography (CT) systems. This study aimed to derive and validate proposed HU-to-density calibration curves (PCTs) from multi-vendor diagnostic CT (dCT) data, independent of scanner and acquisition parameters.<i>Approach.</i>Both HU-to-relative electron density (RED) and mass density (MD) calibration curves were generated using the CIRS 062 M phantom scanned on eight multi-vendor CT systems at 80-140 kilovoltage-peak (kV<sub>p</sub>), including dual-energy modes. PCT (PCT-RED, PCT-MD) were derived by piecewise regression across the multi-kV<sub>p</sub>dataset. 20 palliative 6 MV flattening-filter-free (FFF) volumetric modulated arc therapy and intensity-modulated RT plans were recalculated in Eclipse v18 using all these curves with Acuros XB (dose-to-medium, MD-based) and the anisotropic analytical algorithm (RED-based). Dosimetric comparisons assessed: (1) PCT validation versus scanner-kV<sub>p</sub>-specific calibrations; (2) intra-scanner effects of varying kV<sub>p</sub>(versus 120/110 kV<sub>p</sub>per scanner). Evaluation included dose-volume histogram (DVH) metrics-planning target volume, organs at risk (OARs), body maximum dose, conformity index-and 3D<i>γ</i>analysis (2%/2 mm, 1%/1 mm) between calculated dose distributions. Statistical differences were assessed using the Wilcoxon signed-rank test (<i>α</i>= 0.05) with effect sizes.<i>Main results.</i>Differences between plan dose distributions obtained with PCT and those from scanner-specific curves were minimal. Although several DVH metrics were statistically different (<i>p</i>< 0.05), all differences remained within clinical acceptability. Across both PCT validation and intra-scanner kV<sub>p</sub>variation scenarios, median dosimetric differences were < 1.0% for both targets and OARs, regardless of HU variability. All plans achieved<i>γ</i>passing rates > 99% (1%/1 mm), confirming robustness.<i>Significance.</i>We propose a CT calibration model based on dCT for RT. This multi-vendor approach provides default curves when scanner-specific calibration is impractical, facilitating resource-efficient, simulation-free RT workflows via dCT-based treatment planning, while maintaining dosimetric accuracy across the evaluated dataset.
Authors
- Antonio Javier García Sánchez
- Jonathan Suárez Arteaga
- David Ramos Amores
- Miguel Martínez Albaladejo
- Alfredo Serna
- Francisco Martínez Sánchez
Institutions
- Universidad Politécnica de Cartagena (ES)
- Santa Lucía University General Hospital (ES)
- Universidad de Murcia (ES)
Publication Details
- Journal
- Physics in Medicine and Biology
- Published
- 2026-02-16
- DOI
- https://doi.org/10.1088/1361-6560/ae4668
- Primary Topic
- Advanced Radiotherapy Techniques
- Type
- article
- Field-Weighted Citation Impact
- 0.00