Integrated monte carlo simulation and experimental approach for efficiency calibration of an HPGe detector using bulk environmental samples
Calibration of high-purity germanium (HPGe) detectors for bulk environmental samples is particularly challenging due to complex geometries, heterogeneous matrices, and coincidence summing effects. In this study, we address these issues by validating a detailed MCNPX model of an N-type HPGe detector against experimental measurements performed with a certified multi-gamma source (12ML01) encapsulated in a 0.45 L Marinelli beaker. The comparison between experimental full-energy peak efficiencies and simulations shows strong agreement, with deviations within 1.2% across the energy range of 59 to 1836 keV. These discrepancies are mainly linked to coincidence summing and uncertainties in dead-layer characterization. The validated model provides a reliable and flexible calibration strategy that extends beyond the limited lifetime of physical standards, offering a practical and cost-effective solution for routine environmental radioactivity monitoring.
Authors
- Tarek Azli (ORCID: https://orcid.org/0000-0003-2011-7487)
- Samir Mazidi
Institutions
- Centre de Recherche Nucléaire d’Alger (DZ)
Publication Details
- Journal
- Radiation Protection Dosimetry
- Published
- 2025-10-28
- DOI
- https://doi.org/10.1093/rpd/ncaf138
- Primary Topic
- Radioactivity and Radon Measurements
- Type
- article
- Field-Weighted Citation Impact
- 0.00