GC-FID Quantification of Methanol, Ethanol, 2-Propanol, Acetone, Ethylene Glycol, Diethylene Glycol, 1,2-Propylene Glycol and 1,3-Propylene Glycol in Plasma

Alcohol and glycol ingestion, including substances such as methanol, ethylene glycol, and isopropanol, constitutes a serious medical emergency that requires prompt diagnosis and treatment of patients. Rapid and accurate quantification of these compounds in plasma is essential to guide clinical decision-making and prevent delays in treatment that could result in irreversible organ damage or death. Many healthcare facilities lack analytical methods capable of simultaneously quantifying both alcohols and glycols in a single run. This study describes the development and validation of a rapid gas chromatography-flame ionization detection (GC-FID) method for the simultaneous screening and quantification of methanol, ethanol, 2-propanol, acetone, ethylene glycol, diethylene glycol, 1,2-propylene glycol and 1,3-propylene glycol in human plasma. Plasma samples were prepared using a protein precipitation technique with acetonitrile containing two internal standards: 2-butanol and 1,4-butanediol. Acetonitrile effectively precipitated plasma proteins. The supernatant was then subjected to GC-FID analysis for quantification of the target alcohols and glycols. The total analytical run time was 5 minutes, enabling the quantification of eight analytes in a single injection. The method demonstrated excellent linearity, with correlation coefficients (R2) exceeding 0.9995 for all compounds. The linear dynamic range was 40-1280 mg/L for methanol, 2-propanol, acetone, ethylene glycol, diethylene glycol, 1,2-propylene glycol, and 1,3-propylene glycol, and 80-2560 mg/L for ethanol. Within-run and between-run precision and accuracy (CV and bias) for all analytes were within the accepted criteria of ± 15%. No significant interference, carry-over, or matrix effects were observed, confirming the method's selectivity and robustness. The developed GC-FID method enables rapid, accurate, and simultaneous quantification of toxic alcohols and glycols in plasma within a 5-minute run time. The excellent linearity, precision, and selectivity of the method met all analytical performance criteria, making it well-suited for routine clinical use. This method provides a valuable tool for timely diagnosis and management of suspected toxic alcohol and glycol ingestions in patients in emergency settings.

Authors

Institutions

Publication Details

Journal
Journal of Analytical Toxicology
Published
2026-02-09
DOI
https://doi.org/10.1093/jat/bkag013
Primary Topic
Poisoning and overdose treatments
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

GC-FID Quantification of Methanol, Ethanol, 2-Propanol, Acetone, Ethylene Glycol, Diethylene Glycol, 1,2-Propylene Glycol and 1,3-Propylene Glycol in Plasma

Kim C. M. van der Elst, Tim Bognàr, Erin H. Smeijsters, Mohsin El Amrani et al.
Journal of Analytical Toxicology
Poisoning and overdose treatments
article

GC-FID Quantification of Methanol, Ethanol, 2-Propanol, Acetone, Ethylene Glycol, Diethylene Glycol, 1,2-Propylene Glycol and 1,3-Propylene Glycol in Plasma

Kim C. M. van der Elst, Tim Bognàr, Erin H. Smeijsters, Mohsin El Amrani, Arjan Zandbergen, Mike Groot
article en

Abstract

Alcohol and glycol ingestion, including substances such as methanol, ethylene glycol, and isopropanol, constitutes a serious medical emergency that requires prompt diagnosis and treatment of patients. Rapid and accurate quantification of these compounds in plasma is essential to guide clinical decision-making and prevent delays in treatment that could result in irreversible organ damage or death. Many healthcare facilities lack analytical methods capable of simultaneously quantifying both alcohols and glycols in a single run. This study describes the development and validation of a rapid gas chromatography-flame ionization detection (GC-FID) method for the simultaneous screening and quantification of methanol, ethanol, 2-propanol, acetone, ethylene glycol, diethylene glycol, 1,2-propylene glycol and 1,3-propylene glycol in human plasma. Plasma samples were prepared using a protein precipitation technique with acetonitrile containing two internal standards: 2-butanol and 1,4-butanediol. Acetonitrile effectively precipitated plasma proteins. The supernatant was then subjected to GC-FID analysis for quantification of the target alcohols and glycols. The total analytical run time was 5 minutes, enabling the quantification of eight analytes in a single injection. The method demonstrated excellent linearity, with correlation coefficients (R2) exceeding 0.9995 for all compounds. The linear dynamic range was 40-1280 mg/L for methanol, 2-propanol, acetone, ethylene glycol, diethylene glycol, 1,2-propylene glycol, and 1,3-propylene glycol, and 80-2560 mg/L for ethanol. Within-run and between-run precision and accuracy (CV and bias) for all analytes were within the accepted criteria of ± 15%. No significant interference, carry-over, or matrix effects were observed, confirming the method's selectivity and robustness. The developed GC-FID method enables rapid, accurate, and simultaneous quantification of toxic alcohols and glycols in plasma within a 5-minute run time. The excellent linearity, precision, and selectivity of the method met all analytical performance criteria, making it well-suited for routine clinical use. This method provides a valuable tool for timely diagnosis and management of suspected toxic alcohol and glycol ingestions in patients in emergency settings.

Journal of Analytical Toxicology
Utrecht University (NL), University Medical Center Utrecht (NL)
Peace, Justice and strong institutions
Openalex Percentile: Top 57%
Poisoning and overdose treatments
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.