Microdroplet digestion of proteins, including monoclonal antibodies (mAbs), has gained increasing attention for its potential in rapid protein structural characterization. In this study, we developed a fast antibody characterization and quantitation method using online microdroplet trypsin digestion. This automated approach involves ultrafast digestion in <1 ms (>90% digestion efficiency) and subsequent MS and MS/MS for peptide mapping and pinpointing peptide modifications. For instance, by MS/MS analysis, the N<sup>387</sup> residue rather than N<sup>392</sup> and N<sup>393</sup> of the "PENNYK" peptide, GFYPSDIAVEWESN<sup>387</sup>GQPEN<sup>392</sup>N<sup>393</sup>YK generated from NIST mAb microdroplet digestion, was found to undergo major deamidation due to its proximity to the G<sup>388</sup> residue (upon incubation with a pH 8 Tris buffer for 5 days, ca. 32% of the N<sup>387</sup> residue was deamidated). To further demonstrate our method's applications, methionine oxidation, another important type of post-translational modification (PTM) of mAbs, was successfully quantified by spiking the mAb sample with standard peptides for microdroplet digestion. Such an absolute quantitation approach showed a better accuracy for measuring the methionine oxidation level, in comparison to traditional relative quantitation methods, which simply compare ion intensities of oxidized and intact peptides to calculate the oxidation level (the relative quantitation is problematic as the oxidized and intact peptides have different ionization efficiencies). Similarly, by spiking the antibody with a heavy isotope-labeled antibody, our microdroplet digestion method allowed quick absolute antibody quantification, demonstrating good linearity (<i>R</i><sup>2</sup> = 0.99), sensitivity (LOD: 1.2 ng), and accuracy (0.6-10% quantification error in comparison to the theoretical amount of injected antibody).

Authors

Institutions

Publication Details

Journal
Journal of the American Society for Mass Spectrometry
Published
2026-02-15
DOI
https://doi.org/10.1021/jasms.5c00457
Primary Topic
Advanced Proteomics Techniques and Applications
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Rapid Antibody Structural Characterization and Quantification via Microdroplet Trypsin Digestion

Andrew Schladebeck, Harsha P. Gunawardena, Xi Qiu, Karen Yao Luo et al.
Journal of the American Society for Mass Spectrometry
Advanced Proteomics Techniques and Applications
article

Rapid Antibody Structural Characterization and Quantification via Microdroplet Trypsin Digestion

Andrew Schladebeck, Harsha P. Gunawardena, Xi Qiu, Karen Yao Luo, Michael D. Knierman, Hui Zhao, Yongqing Yang, Mengyuan Xiao, Jim Lau, Hao Chen
article en

Abstract

Microdroplet digestion of proteins, including monoclonal antibodies (mAbs), has gained increasing attention for its potential in rapid protein structural characterization. In this study, we developed a fast antibody characterization and quantitation method using online microdroplet trypsin digestion. This automated approach involves ultrafast digestion in <1 ms (>90% digestion efficiency) and subsequent MS and MS/MS for peptide mapping and pinpointing peptide modifications. For instance, by MS/MS analysis, the N<sup>387</sup> residue rather than N<sup>392</sup> and N<sup>393</sup> of the "PENNYK" peptide, GFYPSDIAVEWESN<sup>387</sup>GQPEN<sup>392</sup>N<sup>393</sup>YK generated from NIST mAb microdroplet digestion, was found to undergo major deamidation due to its proximity to the G<sup>388</sup> residue (upon incubation with a pH 8 Tris buffer for 5 days, ca. 32% of the N<sup>387</sup> residue was deamidated). To further demonstrate our method's applications, methionine oxidation, another important type of post-translational modification (PTM) of mAbs, was successfully quantified by spiking the mAb sample with standard peptides for microdroplet digestion. Such an absolute quantitation approach showed a better accuracy for measuring the methionine oxidation level, in comparison to traditional relative quantitation methods, which simply compare ion intensities of oxidized and intact peptides to calculate the oxidation level (the relative quantitation is problematic as the oxidized and intact peptides have different ionization efficiencies). Similarly, by spiking the antibody with a heavy isotope-labeled antibody, our microdroplet digestion method allowed quick absolute antibody quantification, demonstrating good linearity (<i>R</i><sup>2</sup> = 0.99), sensitivity (LOD: 1.2 ng), and accuracy (0.6-10% quantification error in comparison to the theoretical amount of injected antibody).

Journal of the American Society for Mass Spectrometry
New Jersey Institute of Technology (US), Johnson & Johnson (United States) (US), Agilent Technologies (United States) (US)
Openalex Percentile: Top 56%
Advanced Proteomics Techniques and Applications
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.