Measurement of Scattered Radiation Using Ionization Chamber-Based Survey Meter with Variable Phantom Thickness and Kilovoltage in a Digital Radiographic Room

Abstract Scattered radiation in diagnostic radiology is a source of radiation exposure to medical staff. It mostly originates from patient interaction during X-ray exposure. Its intensity is determined by technical conditions, including kilovoltage (kVp), phantom thickness, and field size. This research endeavors to quantify scatter radiation with an ionization chamber-based survey meter for different conditions of kVp and phantom thickness to assess its effect in a digital radiographic room. A GE Definium XR-8000 digital X-ray machine served as the source of radiation. Scattered radiation at a standard distance and angle was measured with the help of the Fluke 451P ionization chamber-based survey meter. Both a basic water phantom and a general radiographic polymethyl methacrylate phantom were employed with varying thicknesses (5–25 cm and 10–40 mm, respectively). Scattered radiation was measured across different tube voltages (50–125 kVp), while mAs and source-to-image distance were kept constant. Scattered radiation was proportional to both tube voltage and phantom thickness. In 25-cm water phantom thickness, scatter went from 0.38 mR at 50 kVp to 2.05 mR at 125 kVp. A rise in water phantom thickness from 5 to 25 cm at 80 kVp revealed a rise in scatter from 1.67 to 1.9 mR. The radiographic phantom exhibited a similar trend. Present values were higher as compared with values in literature, because the mAs setting was higher for this investigation. This study shows increased scatter with higher kVp and phantom thickness, emphasizing optimized parameters and ALARA (As Low As Reasonably Achievable) for radiation safety.

Authors

Institutions

Publication Details

Journal
Indian journal of radiology and imaging - new series/Indian journal of radiology and imaging/Indian Journal of Radiology & Imaging
Published
2026-02-14
DOI
https://doi.org/10.1055/s-0046-1815933
Primary Topic
Radiation Dose and Imaging
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Measurement of Scattered Radiation Using Ionization Chamber-Based Survey Meter with Variable Phantom Thickness and Kilovoltage in a Digital Radiographic Room

AK Srivastava, Santosh Yadav, Atul Mishra, Rohit Sharma et al.
Indian journal of radiology and imaging - new series/Indian journal of radiology and imaging/Indian Journal of Radiology & Imaging
Radiation Dose and Imaging
article

Measurement of Scattered Radiation Using Ionization Chamber-Based Survey Meter with Variable Phantom Thickness and Kilovoltage in a Digital Radiographic Room

AK Srivastava, Santosh Yadav, Atul Mishra, Rohit Sharma, Abhishek Kumar
article en

Abstract

Abstract Scattered radiation in diagnostic radiology is a source of radiation exposure to medical staff. It mostly originates from patient interaction during X-ray exposure. Its intensity is determined by technical conditions, including kilovoltage (kVp), phantom thickness, and field size. This research endeavors to quantify scatter radiation with an ionization chamber-based survey meter for different conditions of kVp and phantom thickness to assess its effect in a digital radiographic room. A GE Definium XR-8000 digital X-ray machine served as the source of radiation. Scattered radiation at a standard distance and angle was measured with the help of the Fluke 451P ionization chamber-based survey meter. Both a basic water phantom and a general radiographic polymethyl methacrylate phantom were employed with varying thicknesses (5–25 cm and 10–40 mm, respectively). Scattered radiation was measured across different tube voltages (50–125 kVp), while mAs and source-to-image distance were kept constant. Scattered radiation was proportional to both tube voltage and phantom thickness. In 25-cm water phantom thickness, scatter went from 0.38 mR at 50 kVp to 2.05 mR at 125 kVp. A rise in water phantom thickness from 5 to 25 cm at 80 kVp revealed a rise in scatter from 1.67 to 1.9 mR. The radiographic phantom exhibited a similar trend. Present values were higher as compared with values in literature, because the mAs setting was higher for this investigation. This study shows increased scatter with higher kVp and phantom thickness, emphasizing optimized parameters and ALARA (As Low As Reasonably Achievable) for radiation safety.

Indian journal of radiology and imaging - new series/Indian journal of radiology and imaging/Indian Journal of Radiology & Imaging
University College of Medical Sciences (IN), Chhatrapati Shahu Ji Maharaj University (IN), Sarojini Naidu Medical College (IN), Guru Teg Bahadur Hospital (IN), Adesh University (IN), All India Institute of Medical Sciences (IN)
Openalex Percentile: Top 55%
Radiation Dose and Imaging
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.