Interference Corrections During EDXRF Determination of Molybdenum and Sulphur in Powder Samples

ABSTRACT Complete elemental characterization of Molybdenum sulfide, an electro‐catalyst for hydrogen evolution reactions, is essential without much compromise on the material integrity. Energy dispersive x‐ray fluorescence (EDXRF) spectrometry can non‐destructively determine the chemical composition. Mo and S were determined by EDXRF, after addressing the matrix interferences. The interferences due to attenuation of characteristic low energy x‐rays of S by the Mo‐matrix and overlap of S and Mo characteristic x‐rays, lead to rotational and translational matrix effects. These effects on calibration curves were corrected by adopting multiple standard additions, along with matrix correction factors. Herein, the pragmatic inferences, obtained via optimization of experimental variables, were combined with a theoretically calculated correction factor, to solve the extreme spectral overlap problem, during S determinations in the presence of Mo. The results were validated using QC standards and the methodology was applied for real sample analyses. Accuracies better than ±5% were achieved.

Authors

Institutions

Publication Details

Journal
ChemistrySelect
Published
2026-02-01
DOI
https://doi.org/10.1002/slct.202506850
Primary Topic
X-ray Spectroscopy and Fluorescence Analysis
Type
article
Field-Weighted Citation Impact
0.00

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Interference Corrections During EDXRF Determination of Molybdenum and Sulphur in Powder Samples

Nicy Ajith, Remya Devi P.S., Trupti A. Chavan
ChemistrySelect
X-ray Spectroscopy and Fluorescence Analysis
article

Interference Corrections During EDXRF Determination of Molybdenum and Sulphur in Powder Samples

Nicy Ajith, Remya Devi P.S., Trupti A. Chavan
article en

Abstract

ABSTRACT Complete elemental characterization of Molybdenum sulfide, an electro‐catalyst for hydrogen evolution reactions, is essential without much compromise on the material integrity. Energy dispersive x‐ray fluorescence (EDXRF) spectrometry can non‐destructively determine the chemical composition. Mo and S were determined by EDXRF, after addressing the matrix interferences. The interferences due to attenuation of characteristic low energy x‐rays of S by the Mo‐matrix and overlap of S and Mo characteristic x‐rays, lead to rotational and translational matrix effects. These effects on calibration curves were corrected by adopting multiple standard additions, along with matrix correction factors. Herein, the pragmatic inferences, obtained via optimization of experimental variables, were combined with a theoretically calculated correction factor, to solve the extreme spectral overlap problem, during S determinations in the presence of Mo. The results were validated using QC standards and the methodology was applied for real sample analyses. Accuracies better than ±5% were achieved.

ChemistrySelectVol. 11(7)
Bhabha Atomic Research Centre (IN), Homi Bhabha National Institute (IN), Bhabha Atomic Research Center Hospital (IN)
Bhabha Atomic Research Centre
Openalex Percentile: Top 50%
X-ray Spectroscopy and Fluorescence Analysis
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.