<bold>Background</bold>Neutron dose measurement faces challenges due to the wide range of neutron energy and the complexity of energy response. The response of electronic neutron personal dosimeters (n-EPDs) varies significantly under different energy spectra, and calibration with isotope neutron sources may lead to on-site dose measurement deviations.<bold>Purpose</bold>This study aims to investigate the influences of neutron energy spectrum distribution and incident direction on the response of n-EPDs.<bold>Methods</bold>Firstly, <sup>252</sup>Cf isotope neutron source, D-T accelerator neutron source and materials such as heavy water, tungsten, iron, graphite and polyethylene, were used to establish four neutron radiation fields with different energy spectra. Then, the Bonner neutron spectrometer was employed to measure neutron energy spectra, while the incident angle distributions were calculated using Monte Carlo method. Finally, the response differences of two commercial electric personal neutron dosimeters (n-EPDs) were experimentally tested.<bold>Results</bold>The relative deviation between measured spectrum and standard spectrum of <sup>252</sup>Cf reference radiation field is within ±10% in the main energy regions. For four neutron fields, the fluence-averaged neutron energy ranges from 0.12 MeV to 2.00 MeV, the dose-averaged neutron energy ranges from 0.75 MeV to 3.65 MeV, and the spectra-averaged conversion coefficient ranges from 51.17 pSv·cm<sup>2</sup> to 390.3 pSv·cm<sup>2</sup>. The maximum response differences of two commercial n-EPDs to four neutron fields are 5.7 and 6.7 times, respectively.<bold>Conclusions</bold>Results of this study indicate that when using isotope neutron sources for calibration, n-EPDs readings often overestimate the actual individual dose equivalent in workplace, hence special attentions need to be paid to the impacts of energy response, neutron energy spectrum, and incident direction on measurement accuracy.

Authors

Publication Details

Journal
Nuclear Techniques
Published
2026-02-01
DOI
https://doi.org/10.3724/j.0253-3219.2026.hjs.49.250006
Primary Topic
Radiation Therapy and Dosimetry
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Experimental study on neutron response of electronic personal dosimeters based on simulated workplace neutron field

Hongtao Chen, Zhenglin Huang, Zhihui Tang, Liye Liu et al.
Nuclear Techniques
Radiation Therapy and Dosimetry
article

Experimental study on neutron response of electronic personal dosimeters based on simulated workplace neutron field

Hongtao Chen, Zhenglin Huang, Zhihui Tang, Liye Liu, Faguo CHEN, Deyuan LI, Kai Zhang, Hui Li
article en

Abstract

<bold>Background</bold>Neutron dose measurement faces challenges due to the wide range of neutron energy and the complexity of energy response. The response of electronic neutron personal dosimeters (n-EPDs) varies significantly under different energy spectra, and calibration with isotope neutron sources may lead to on-site dose measurement deviations.<bold>Purpose</bold>This study aims to investigate the influences of neutron energy spectrum distribution and incident direction on the response of n-EPDs.<bold>Methods</bold>Firstly, <sup>252</sup>Cf isotope neutron source, D-T accelerator neutron source and materials such as heavy water, tungsten, iron, graphite and polyethylene, were used to establish four neutron radiation fields with different energy spectra. Then, the Bonner neutron spectrometer was employed to measure neutron energy spectra, while the incident angle distributions were calculated using Monte Carlo method. Finally, the response differences of two commercial electric personal neutron dosimeters (n-EPDs) were experimentally tested.<bold>Results</bold>The relative deviation between measured spectrum and standard spectrum of <sup>252</sup>Cf reference radiation field is within ±10% in the main energy regions. For four neutron fields, the fluence-averaged neutron energy ranges from 0.12 MeV to 2.00 MeV, the dose-averaged neutron energy ranges from 0.75 MeV to 3.65 MeV, and the spectra-averaged conversion coefficient ranges from 51.17 pSv·cm<sup>2</sup> to 390.3 pSv·cm<sup>2</sup>. The maximum response differences of two commercial n-EPDs to four neutron fields are 5.7 and 6.7 times, respectively.<bold>Conclusions</bold>Results of this study indicate that when using isotope neutron sources for calibration, n-EPDs readings often overestimate the actual individual dose equivalent in workplace, hence special attentions need to be paid to the impacts of energy response, neutron energy spectrum, and incident direction on measurement accuracy.

Nuclear TechniquesVol. 49(2)
Affordable and clean energy
Openalex Percentile: Top 57%
Radiation Therapy and Dosimetry
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.