To evaluate cross-platform measurement consistency and diagnostic threshold requirements in shear wave elastography (SWE), this study presents a robotically controlled, phantom-based validation framework to quantify and interpret inter-vendor variability that limits clinical standardisation. A custom-fabricated polyvinyl chloride-graphite phantom containing eight spherical inclusions (15-25 mm diameter, 25-95 mm depth, 23.53-259.58 kPa stiffness), representing breast tissue mechanical properties, was evaluated using Samsung HS50 and Aixplorer ultrasound systems. Robotic automation standardised probe positioning and contact, eliminating operator-dependent variability and enabling direct, system-level comparison. Cross-platform reproducibility, accuracy against mechanically validated ground truth, and diagnostic threshold performance were assessed across 80 measurements. Both systems demonstrated excellent intra-machine reproducibility (coefficient of variation: Samsung 0.42%, Aixplorer 0.46%) with strong inter-machine correlation (r = 0.9951, <i>p</i> < 0.0001). However, systematic bias of 7.05 kPa and 95% limits of agreement spanning 38.7 kPa revealed substantial cross-platform measurement differences. All phantom inclusions (8/8) measured below their assigned ground truth stiffness on both systems, with systematic underestimation ranging from 0.33 kPa to 109.57 kPa, indicating parameter-dependent measurement distortion linked to inclusion size, depth, and stiffness. Dynamic range compression was observed (Samsung: 68.7%, Aixplorer: 59.1% of true phantom range), providing a mechanistic explanation for diagnostic threshold instability. This study contributes an interpretable validation methodology that links SWE measurement bias to physical lesion properties and imaging system characteristics, rather than relying on correlation alone. Despite strong reproducibility, the observed system-dependent bias demonstrates that SWE measurements are not directly transferable across ultrasound platforms, and system-specific collaboration is required to ensure reliable clinical interpretation.

Authors

Institutions

Publication Details

Journal
Bioengineering
Published
2026-02-13
DOI
https://doi.org/10.3390/bioengineering13020214
Citations
1
Primary Topic
Ultrasound Imaging and Elastography
Type
article
Field-Weighted Citation Impact
31.92
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Quantitative Shear Wave Elastography: A Phantom—Based Comparison of Two Ultrasound Systems

Sarah Savaridas, Wadhhah Aldehani, Luigi Manfredi
1 citations
Bioengineering
Ultrasound Imaging and Elastography
31.92
article

Quantitative Shear Wave Elastography: A Phantom—Based Comparison of Two Ultrasound Systems

Sarah Savaridas, Wadhhah Aldehani, Luigi Manfredi
article en
1 citations

Abstract

To evaluate cross-platform measurement consistency and diagnostic threshold requirements in shear wave elastography (SWE), this study presents a robotically controlled, phantom-based validation framework to quantify and interpret inter-vendor variability that limits clinical standardisation. A custom-fabricated polyvinyl chloride-graphite phantom containing eight spherical inclusions (15-25 mm diameter, 25-95 mm depth, 23.53-259.58 kPa stiffness), representing breast tissue mechanical properties, was evaluated using Samsung HS50 and Aixplorer ultrasound systems. Robotic automation standardised probe positioning and contact, eliminating operator-dependent variability and enabling direct, system-level comparison. Cross-platform reproducibility, accuracy against mechanically validated ground truth, and diagnostic threshold performance were assessed across 80 measurements. Both systems demonstrated excellent intra-machine reproducibility (coefficient of variation: Samsung 0.42%, Aixplorer 0.46%) with strong inter-machine correlation (r = 0.9951, <i>p</i> < 0.0001). However, systematic bias of 7.05 kPa and 95% limits of agreement spanning 38.7 kPa revealed substantial cross-platform measurement differences. All phantom inclusions (8/8) measured below their assigned ground truth stiffness on both systems, with systematic underestimation ranging from 0.33 kPa to 109.57 kPa, indicating parameter-dependent measurement distortion linked to inclusion size, depth, and stiffness. Dynamic range compression was observed (Samsung: 68.7%, Aixplorer: 59.1% of true phantom range), providing a mechanistic explanation for diagnostic threshold instability. This study contributes an interpretable validation methodology that links SWE measurement bias to physical lesion properties and imaging system characteristics, rather than relying on correlation alone. Despite strong reproducibility, the observed system-dependent bias demonstrates that SWE measurements are not directly transferable across ultrasound platforms, and system-specific collaboration is required to ensure reliable clinical interpretation.

BioengineeringVol. 13(2)
University of Dundee (GB), Ninewells Hospital (GB), Al-Sabah Hospital (KW), Genomics (United Kingdom) (GB)
Openalex Percentile: Top 1%
Ultrasound Imaging and Elastography
31.92
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.