<i>Objective.</i>High-resolution small-animal single-photon emission computed tomography (SPECT) systems require accurate system matrices for high-quality image reconstruction. Experimentally measured point-source system responses provide high fidelity but are time-consuming, and the short half-life of<sup>99m</sup>Tc limits long-duration acquisitions required for dense system-matrix sampling.<i>Approach.</i>We propose a hybrid ⁹⁹Mo/<sup>99m</sup>Tc point source based on ⁹⁹Mo-adsorbable ion-exchange resin microspheres to achieve high activity with an effectively extended usable half-life. High-activity ⁹⁹Mo was adsorbed onto sub-millimeter resin beads, and Monte Carlo simulations were performed to quantitatively evaluate the impact of ⁹⁹Mo-derived high-energy photons on<sup>99m</sup>Tc system-matrix acquisition for two representative high-resolution SPECT configurations: a multi-pinhole system and a self-collimating system.<i>Main results.</i>Resin microspheres with activities exceeding 10 mCi per bead were successfully fabricated. Simulation results demonstrate that, within the<sup>99m</sup>Tc photopeak energy window, the contribution of ⁹⁹Mo-derived high-energy photons produces minimal distortion of system-response centroids and spatial distributions. The proposed source supports extended acquisitions for dense system-matrix sampling (e.g. 100 × 100 × 100 positions) with significantly improved statistical stability compared to conventional<sup>99m</sup>Tc sources.<i>Significance.</i>This work provides a practical and reproducible solution for accurate system-matrix calibration in state-of-the-art high-resolution SPECT systems, particularly where long acquisition durations are required, facilitating consistent system characterization in preclinical imaging research.

Authors

Institutions

Publication Details

Journal
Physics in Medicine and Biology
Published
2026-02-13
DOI
https://doi.org/10.1088/1361-6560/ae45e5
Primary Topic
Medical Imaging Techniques and Applications
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

High-activity ⁹⁹Mo microresin fabrication for submillimeter SPECT system matrix acquisition

Tiantian Dai, Nianming Jiang, Zhenlei Lyu, Debin Zhang et al.
Physics in Medicine and Biology
Medical Imaging Techniques and Applications
article

High-activity ⁹⁹Mo microresin fabrication for submillimeter SPECT system matrix acquisition

Tiantian Dai, Nianming Jiang, Zhenlei Lyu, Debin Zhang, Qingyang Wei, Yuhang Qiu, Tianyu Ma, Lifeng Sun
article en

Abstract

<i>Objective.</i>High-resolution small-animal single-photon emission computed tomography (SPECT) systems require accurate system matrices for high-quality image reconstruction. Experimentally measured point-source system responses provide high fidelity but are time-consuming, and the short half-life of<sup>99m</sup>Tc limits long-duration acquisitions required for dense system-matrix sampling.<i>Approach.</i>We propose a hybrid ⁹⁹Mo/<sup>99m</sup>Tc point source based on ⁹⁹Mo-adsorbable ion-exchange resin microspheres to achieve high activity with an effectively extended usable half-life. High-activity ⁹⁹Mo was adsorbed onto sub-millimeter resin beads, and Monte Carlo simulations were performed to quantitatively evaluate the impact of ⁹⁹Mo-derived high-energy photons on<sup>99m</sup>Tc system-matrix acquisition for two representative high-resolution SPECT configurations: a multi-pinhole system and a self-collimating system.<i>Main results.</i>Resin microspheres with activities exceeding 10 mCi per bead were successfully fabricated. Simulation results demonstrate that, within the<sup>99m</sup>Tc photopeak energy window, the contribution of ⁹⁹Mo-derived high-energy photons produces minimal distortion of system-response centroids and spatial distributions. The proposed source supports extended acquisitions for dense system-matrix sampling (e.g. 100 × 100 × 100 positions) with significantly improved statistical stability compared to conventional<sup>99m</sup>Tc sources.<i>Significance.</i>This work provides a practical and reproducible solution for accurate system-matrix calibration in state-of-the-art high-resolution SPECT systems, particularly where long acquisition durations are required, facilitating consistent system characterization in preclinical imaging research.

Physics in Medicine and BiologyVol. 71(4)
Chengdu Medical College (CN), China-Japan Friendship Hospital (CN), University of Science and Technology Beijing (CN), Tsinghua University (CN)
Affordable and clean energy
Openalex Percentile: Top 57%
Medical Imaging Techniques and Applications
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.