Toward Enhanced Selection of Antibiotics in a Hospital Laboratory—Development of a Rapid Electrochemical Antimicrobial Sensitivity Test for Urinary Tract Infections

Antimicrobial resistance poses a major global health threat, often driven by empirical antibiotic use due to the relatively slow (24-72 h) diagnostic pathway for urinary tract infections (UTIs). Due to their low cost and ease of multiplexing, electrochemical biosensors offer a promising alternative. Herein, we report on the Microplate Dx RapidPlate platform, a rapid electrochemical impedance spectroscopy (EIS)-based approach for accelerated antibiotic susceptibility testing (AST) directly from urine samples, without pre-treatment. This report charts the evolution of the core EIS-based AST measurement from a laboratory prototype into a fully realized <i>Alpha</i> prototype platform with an accompanying 32-channel electrochemical cartridge designed specifically for interrogation of urine samples. A total of 89 clinically derived urine samples were evaluated at three clinical sites (U.K. and India), with results compared against standard methods (VITEK 2 and disk diffusion). The impedance-based RapidPlate system detected bacterial growth within 50 min and provided susceptibility profiles by 85 min, with 95% concordance to reference tests. Consistent performance was demonstrated across diverse clinical settings and organisms, including <i>Escherichia coli</i> and <i>Klebsiella pneumoniae</i>. The RapidPlate AST system has demonstrated great potential as a cost-effective, point-of-care tool for rapid AST, supporting and improving UTI management and antimicrobial stewardship.

Authors

Institutions

Publication Details

Journal
ACS Sensors
Published
2026-02-13
DOI
https://doi.org/10.1021/acssensors.5c02725
Primary Topic
Bacterial Identification and Susceptibility Testing
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Toward Enhanced Selection of Antibiotics in a Hospital Laboratory—Development of a Rapid Electrochemical Antimicrobial Sensitivity Test for Urinary Tract Infections

A. Poojary, Michael Malecha, Alistair Longmuir, Vincent Vezza et al.
ACS Sensors
Bacterial Identification and Susceptibility Testing
article

Toward Enhanced Selection of Antibiotics in a Hospital Laboratory—Development of a Rapid Electrochemical Antimicrobial Sensitivity Test for Urinary Tract Infections

A. Poojary, Michael Malecha, Alistair Longmuir, Vincent Vezza, Damion K. Corrigan, Heather Ralston, Adrian Butterworth, Paul A. Hoskisson, Ben Clark, Amanda Winter, Michael E. Murphy, Emma C. L. Marrs, Maria Katsafadou, David Price, Stuart Hannah, Tyla Campbell
article en

Abstract

Antimicrobial resistance poses a major global health threat, often driven by empirical antibiotic use due to the relatively slow (24-72 h) diagnostic pathway for urinary tract infections (UTIs). Due to their low cost and ease of multiplexing, electrochemical biosensors offer a promising alternative. Herein, we report on the Microplate Dx RapidPlate platform, a rapid electrochemical impedance spectroscopy (EIS)-based approach for accelerated antibiotic susceptibility testing (AST) directly from urine samples, without pre-treatment. This report charts the evolution of the core EIS-based AST measurement from a laboratory prototype into a fully realized <i>Alpha</i> prototype platform with an accompanying 32-channel electrochemical cartridge designed specifically for interrogation of urine samples. A total of 89 clinically derived urine samples were evaluated at three clinical sites (U.K. and India), with results compared against standard methods (VITEK 2 and disk diffusion). The impedance-based RapidPlate system detected bacterial growth within 50 min and provided susceptibility profiles by 85 min, with 95% concordance to reference tests. Consistent performance was demonstrated across diverse clinical settings and organisms, including <i>Escherichia coli</i> and <i>Klebsiella pneumoniae</i>. The RapidPlate AST system has demonstrated great potential as a cost-effective, point-of-care tool for rapid AST, supporting and improving UTI management and antimicrobial stewardship.

ACS SensorsVol. 11(2)
University of Strathclyde (GB), Glasgow Royal Infirmary (GB), Breach Candy Hospital (IN), Newcastle upon Tyne Hospitals NHS Foundation Trust (GB), Micro Materials (United Kingdom) (GB)
Openalex Percentile: Top 55%
Bacterial Identification and Susceptibility Testing
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.