Evaluation of Redundancy Methods and Strategies for the On-Board Spectral Reference of the NOAA CrIS Instruments

The cross-track infrared sounder (CrIS) is an infrared Fourier transform spectrometer (FTS) hosted on the Joint Polar Satellite System (JPSS). To produce a CrIS sensor data record (SDR) with high-quality spectral calibration, it is necessary to accurately and periodically calibrate CrIS interferometer metrology laser wavelength over time (i.e., a spectral calibration reference). The current spectral calibration reference method makes use of a neon lamp’s highly stable gas emission spectral line processed through the FTS. Further improvements in spectral calibration reliability have been developed using alternative methods described in this article which provide redundancy to the in-orbit neon lamp spectral calibration reference. The first proposed redundancy method involves a real-time regression algorithm which uses the known correlation between the laser diode’s output wavelength and the laser diode temperature to provide updated estimates of the metrology laser wavelength. A second proposed method uses simulated atmospheric line absorption and emission radiances as a spectral reference to estimate the spectral changes in the FTS laser metrology wavelength. Starting with NOAA-21 CrIS and for all other future CrIS sensors, the periodicity of calibrating the laser’s wavelength was changed from approximately once per satellite orbit to once every seven orbits to extend the neon lamp’s useful lifetime. Here we present the estimated impact of changing the interval between neon lamp measurements of the CrIS metrology laser wavelength, along with studies of these two redundancy methods and their assessment in an operational environment.

Authors

Publication Details

Journal
Maryland Shared Open Access Repository (USMAI Consortium)
Published
2026-01-14
DOI
https://doi.org/10.13016/m2vs8e-x3v5
Primary Topic
Calibration and Measurement Techniques
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Evaluation of Redundancy Methods and Strategies for the On-Board Spectral Reference of the NOAA CrIS Instruments

Sara K Glass, Peter Beierle, Denis Tremblay, Flavio Iturbide-Sanchez et al.
Maryland Shared Open Access Repository (USMAI Consortium)
Calibration and Measurement Techniques
article

Evaluation of Redundancy Methods and Strategies for the On-Board Spectral Reference of the NOAA CrIS Instruments

Sara K Glass, Peter Beierle, Denis Tremblay, Flavio Iturbide-Sanchez, Joseph P. Predina, David F. Johnson, Kun Zhang, L. Larrabee Strow, Yong Chen
article en

Abstract

The cross-track infrared sounder (CrIS) is an infrared Fourier transform spectrometer (FTS) hosted on the Joint Polar Satellite System (JPSS). To produce a CrIS sensor data record (SDR) with high-quality spectral calibration, it is necessary to accurately and periodically calibrate CrIS interferometer metrology laser wavelength over time (i.e., a spectral calibration reference). The current spectral calibration reference method makes use of a neon lamp’s highly stable gas emission spectral line processed through the FTS. Further improvements in spectral calibration reliability have been developed using alternative methods described in this article which provide redundancy to the in-orbit neon lamp spectral calibration reference. The first proposed redundancy method involves a real-time regression algorithm which uses the known correlation between the laser diode’s output wavelength and the laser diode temperature to provide updated estimates of the metrology laser wavelength. A second proposed method uses simulated atmospheric line absorption and emission radiances as a spectral reference to estimate the spectral changes in the FTS laser metrology wavelength. Starting with NOAA-21 CrIS and for all other future CrIS sensors, the periodicity of calibrating the laser’s wavelength was changed from approximately once per satellite orbit to once every seven orbits to extend the neon lamp’s useful lifetime. Here we present the estimated impact of changing the interval between neon lamp measurements of the CrIS metrology laser wavelength, along with studies of these two redundancy methods and their assessment in an operational environment.

Maryland Shared Open Access Repository (USMAI Consortium)
Openalex Percentile: Top 58%
Calibration and Measurement Techniques
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.