Feasibility of proton daily QA prototype for pencil beam scanning carbon ion beam therapy

The DQA-P prototype demonstrated feasibility for daily QA in PBS carbon ion beam therapy, offering a short overall measurement time of approximately 5 min, efficient data acquisition, and simultaneous multiparameter verification within a single setup. This is a significant improvement on the conventional daily QA procedure, which uses two different detector types for independent dose and spot position measurements. In the standard workflow, both detectors must be positioned individually using the in-room laser system, connected to the readout electronics and irradiated using the daily beam plan. Altogether, this procedure takes around 15-20 min. Although certain limitations remain-particularly in spot size accuracy and range determination-the device showed sufficient sensitivity and stability for daily QA purposes. Its use has the potential to streamline QA procedures, reduce beam time, and enhance workflow efficiency in carbon ion therapy.

Authors

Institutions

Publication Details

Journal
Journal of Applied Clinical Medical Physics
Published
2026-02-01
DOI
https://doi.org/10.1002/acm2.70502
Primary Topic
Radiation Therapy and Dosimetry
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Feasibility of proton daily QA prototype for pencil beam scanning carbon ion beam therapy

Veronika Flatten, K Baumann, Uli Weber, Matthias Witt et al.
Journal of Applied Clinical Medical Physics
Radiation Therapy and Dosimetry
article

Feasibility of proton daily QA prototype for pencil beam scanning carbon ion beam therapy

Veronika Flatten, K Baumann, Uli Weber, Matthias Witt, Klemens Zink, Andreas Schönfeld
article en

Abstract

The DQA-P prototype demonstrated feasibility for daily QA in PBS carbon ion beam therapy, offering a short overall measurement time of approximately 5 min, efficient data acquisition, and simultaneous multiparameter verification within a single setup. This is a significant improvement on the conventional daily QA procedure, which uses two different detector types for independent dose and spot position measurements. In the standard workflow, both detectors must be positioned individually using the in-room laser system, connected to the readout electronics and irradiated using the daily beam plan. Altogether, this procedure takes around 15-20 min. Although certain limitations remain-particularly in spot size accuracy and range determination-the device showed sufficient sensitivity and stability for daily QA purposes. Its use has the potential to streamline QA procedures, reduce beam time, and enhance workflow efficiency in carbon ion therapy.

Journal of Applied Clinical Medical PhysicsVol. 27(2)
Philipps University of Marburg (DE), GSI Helmholtz Centre for Heavy Ion Research (DE), Sun Nuclear (United States) (US), Radiation Protection Institute (UA)
Openalex Percentile: Top 60%
Radiation Therapy and Dosimetry
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.