High-Tc SQUID-based ultra-low field MRI in unshielded environments
A high-performance MRI based on high-Tc SQUID and flux coupling at ultralow magnetic fields is reported in this work. The technique of NMR/MRI was optimized to study MRI of water phantoms and fingers. An SNR of 62 in image contrast was achieved for a water column of 200 μL at magnetic fields of 98 μT. The MRI showed high image contrast and demonstrated a spatial resolution of 2 mm for samples in magnetically unshielded environments. This resolution is comparable to the current state-of-the-art values (typically 2.0–3.0 mm) reported for SQUID-based ULF MRI systems. A clear image of fingers was demonstrated for the first time using high-Tc SQUIDs, which shows promise for biological or clinical applications.
Authors
- Shu Liao (ORCID: https://orcid.org/0009-0002-4668-5204)
- H. Huang
Institutions
- National Taiwan Normal University (TW)
Publication Details
- Journal
- AIP Advances
- Published
- 2026-02-01
- DOI
- https://doi.org/10.1063/9.0001031
- Primary Topic
- Advanced MRI Techniques and Applications
- Type
- article
- Field-Weighted Citation Impact
- 0.00
Funders
- National Taiwan Normal University
- National Science and Technology Council