Evaluation of radiological properties and anisotropy with air channels analysis in 3D-printed flexible lung-mimicking materials for radiotherapy

<i>Objective.</i>Three-dimensional (3D) printing is increasingly used for radiotherapy quality assurance (QA) phantoms, yet, reproducing the structural heterogeneity and radiological properties of lung tissue remains challenging. This study evaluates thermoplastic polyurethane (TPU) for dynamic lung-mimicking phantoms, selected for its flexibility while preserving print-defined geometry, with a focus on radiological equivalence, directional anisotropy, and the detectability of sub-millimetre air channels.<i>Approach.</i>Eleven TPU materials with various colours and Shore hardness (63-82) were printed into gyroid-patterned inserts of varying infill densities. Effective atomic number (<i>Z</i><sub>eff</sub>) and relative electron density (RED) were determined using dual-energy computed tomography (CT). Anisotropy and internal air channels were assessed in five orientations using micro-CT, clinical CT, and flat-panel detector (FPD) imaging, and compared to a voxelised digital model derived from G-code toolpaths.<i>Main results</i>. Measured<i>Z</i><sub>eff</sub>ranged from 6.3 ± 0.6 to 11.1 ± 0.2, with pigment-driven variation up to 66% within identical material categories. Seven materials achieved lung mimicking<i>Z</i><sub>eff</sub>(around 7.55). RED increased with infill and mimicked lung references (0.28-0.43) at moderate infill. The low-infill insert contained 0.7 mm air channels visible in all modalities. The high-infill insert contained 0.3 mm channels, resolved accurately by the micro-CT and FPD but not reliably resolved by clinical CT due to resolution limitations. The digital model indicated diagonal anisotropy, micro-CT and FPD indicated near-isotropy in low and high infill, while CT showed apparent anisotropy due to its resolution limitations.<i>Significance</i>. TPU-based gyroid phantoms are suitable lung-mimicking candidates for radiotherapy QA of imaging and dosimetry. Their periodic air-channels are reliably resolved by high-resolution imaging (micro-CT or FPD), but may be distorted by clinical CT, particularly in resolution-limited orientations, the digital model supports pre-print evaluation of these air-channels. Because undetected internal heterogeneities may affect dose calculation accuracy, high-resolution imaging when available, is preferred for assessing the internal structure of 3D-printed phantoms.

Authors

Institutions

Publication Details

Journal
Physics in Medicine and Biology
Published
2026-02-12
DOI
https://doi.org/10.1088/1361-6560/ae456d
Primary Topic
Advanced Radiotherapy Techniques
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Evaluation of radiological properties and anisotropy with air channels analysis in 3D-printed flexible lung-mimicking materials for radiotherapy

Didier Lustermans, Murillo Bellezzo, Brigitte Reniers, Frank Verhaegen et al.
Physics in Medicine and Biology
Advanced Radiotherapy Techniques
article

Evaluation of radiological properties and anisotropy with air channels analysis in 3D-printed flexible lung-mimicking materials for radiotherapy

Didier Lustermans, Murillo Bellezzo, Brigitte Reniers, Frank Verhaegen, Behzad Rezaeifar, Roua Abdulrahim, Gabriel Paiva Fonseca
article en

Abstract

<i>Objective.</i>Three-dimensional (3D) printing is increasingly used for radiotherapy quality assurance (QA) phantoms, yet, reproducing the structural heterogeneity and radiological properties of lung tissue remains challenging. This study evaluates thermoplastic polyurethane (TPU) for dynamic lung-mimicking phantoms, selected for its flexibility while preserving print-defined geometry, with a focus on radiological equivalence, directional anisotropy, and the detectability of sub-millimetre air channels.<i>Approach.</i>Eleven TPU materials with various colours and Shore hardness (63-82) were printed into gyroid-patterned inserts of varying infill densities. Effective atomic number (<i>Z</i><sub>eff</sub>) and relative electron density (RED) were determined using dual-energy computed tomography (CT). Anisotropy and internal air channels were assessed in five orientations using micro-CT, clinical CT, and flat-panel detector (FPD) imaging, and compared to a voxelised digital model derived from G-code toolpaths.<i>Main results</i>. Measured<i>Z</i><sub>eff</sub>ranged from 6.3 ± 0.6 to 11.1 ± 0.2, with pigment-driven variation up to 66% within identical material categories. Seven materials achieved lung mimicking<i>Z</i><sub>eff</sub>(around 7.55). RED increased with infill and mimicked lung references (0.28-0.43) at moderate infill. The low-infill insert contained 0.7 mm air channels visible in all modalities. The high-infill insert contained 0.3 mm channels, resolved accurately by the micro-CT and FPD but not reliably resolved by clinical CT due to resolution limitations. The digital model indicated diagonal anisotropy, micro-CT and FPD indicated near-isotropy in low and high infill, while CT showed apparent anisotropy due to its resolution limitations.<i>Significance</i>. TPU-based gyroid phantoms are suitable lung-mimicking candidates for radiotherapy QA of imaging and dosimetry. Their periodic air-channels are reliably resolved by high-resolution imaging (micro-CT or FPD), but may be distorted by clinical CT, particularly in resolution-limited orientations, the digital model supports pre-print evaluation of these air-channels. Because undetected internal heterogeneities may affect dose calculation accuracy, high-resolution imaging when available, is preferred for assessing the internal structure of 3D-printed phantoms.

Physics in Medicine and BiologyVol. 71(5)
Maastricht University Medical Centre (NL), Maastro Clinic (NL), Hasselt University (BE)
Life below water
Openalex Percentile: Top 53%
Advanced Radiotherapy Techniques
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.