Portable Electrical Impedance Prescreening for Breast tissue suspicious for malignancy: Model Optimization and Clinical Performance of the Julieta Device in a Multicenter Cross-Sectional Study in Colombia

IntroductionBreast cancer remains a leading cause of cancer mortality despite being potentially curable when detected early, particularly in low- and middle-income countries where access to screening is limited. This is largely driven by operational gaps, including limited access to screening and delays in diagnosis and treatment. JULIETA is a portable bioimpedance spectroscopy device designed to identify electrical tissue patterns associated with potentially malignant findings and to prioritize women for further diagnostic evaluation. This study assessed the performance of a hierarchical algorithm integrated into JULIETA to distinguish findings without malignant potential (BI-RADS 1-2) from those with malignant potential (BI-RADS ≥3), using mammography as the reference standard.MethodsA cross-sectional observational study with prospective data collection was conducted between May and July 2024 in four Colombian cities. Adult women undergoing screening or follow-up mammography were evaluated with JULIETA prior to imaging. Impedance-derived features, breast density estimates, and individual risk scores were used to retrain a hierarchical classifier combining Random Forest and SVM-RBF models, using an 80/20 stratified split and cross-validation.ResultsA total of 1350 women were recruited (mean age 56.5 ± 8.0 years); 67% were BI-RADS 1-2 and 21% BI-RADS 4. After data cleaning, 673 breasts (469 women) were included. The model achieved 73% sensitivity, 76% specificity, 65.5% positive predictive value, and 82.1% negative predictive value.ConclusionJULIETA is a feasible, safe, and reproducible noninvasive bioimpedance pre-screening tool that may enable scalable triage and support earlier detection and improved equity when integrated into public health pathways.

Authors

Institutions

Publication Details

Journal
Technology in Cancer Research & Treatment
Published
2026-02-01
DOI
https://doi.org/10.1177/15330338261422902
Citations
1
Primary Topic
Electrical and Bioimpedance Tomography
Type
article
Field-Weighted Citation Impact
21.93
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Portable Electrical Impedance Prescreening for Breast tissue suspicious for malignancy: Model Optimization and Clinical Performance of the Julieta Device in a Multicenter Cross-Sectional Study in Colombia

María Alejandra Yépez, Valentina Villegas González, María Andrea Negret, David Grajales et al.
1 citations
Technology in Cancer Research & Treatment
Electrical and Bioimpedance Tomography
21.93
article

Portable Electrical Impedance Prescreening for Breast tissue suspicious for malignancy: Model Optimization and Clinical Performance of the Julieta Device in a Multicenter Cross-Sectional Study in Colombia

María Alejandra Yépez, Valentina Villegas González, María Andrea Negret, David Grajales, Sergio Lopez, Valentina Agudelo, Clara Piedrahita
article en
1 citations

Abstract

IntroductionBreast cancer remains a leading cause of cancer mortality despite being potentially curable when detected early, particularly in low- and middle-income countries where access to screening is limited. This is largely driven by operational gaps, including limited access to screening and delays in diagnosis and treatment. JULIETA is a portable bioimpedance spectroscopy device designed to identify electrical tissue patterns associated with potentially malignant findings and to prioritize women for further diagnostic evaluation. This study assessed the performance of a hierarchical algorithm integrated into JULIETA to distinguish findings without malignant potential (BI-RADS 1-2) from those with malignant potential (BI-RADS ≥3), using mammography as the reference standard.MethodsA cross-sectional observational study with prospective data collection was conducted between May and July 2024 in four Colombian cities. Adult women undergoing screening or follow-up mammography were evaluated with JULIETA prior to imaging. Impedance-derived features, breast density estimates, and individual risk scores were used to retrain a hierarchical classifier combining Random Forest and SVM-RBF models, using an 80/20 stratified split and cross-validation.ResultsA total of 1350 women were recruited (mean age 56.5 ± 8.0 years); 67% were BI-RADS 1-2 and 21% BI-RADS 4. After data cleaning, 673 breasts (469 women) were included. The model achieved 73% sensitivity, 76% specificity, 65.5% positive predictive value, and 82.1% negative predictive value.ConclusionJULIETA is a feasible, safe, and reproducible noninvasive bioimpedance pre-screening tool that may enable scalable triage and support earlier detection and improved equity when integrated into public health pathways.

Technology in Cancer Research & TreatmentVol. 25
Pontificia Universidad Javeriana (CO), EIA University (CO), Universidad de Antioquia (CO), Colegio de Estudios Superiores de Administración (CO), Municipality of Medellín (CO), Institut universitaire en santé mentale de Montréal (CA), Universidad de Los Andes (BO), Colciencias (CO), Artificial Intelligence in Medicine (Canada) (CA), University of La Salle (CO), Universidad EAFIT (CO)
No poverty
Openalex Percentile: Top 1%
Electrical and Bioimpedance Tomography
21.93
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.