Reduction of radiation dose to the eye lens during common CT examinations of the head, paranasal sinus, and cervical spine in emergency settings: A phantom study
Implementation of the ODM technique resulted in significant eye lens dose reduction (4.28%-47.97%) across head, PNS, and C-spine CT examinations with no significant differences in image noise, SNR, and CNR compared to both fixed tube current and ATCM techniques. ODM demonstrates potential as a practical dose optimization strategy for routine emergency head and neck CT imaging. Further studies with subjective image quality assessment are recommended to evaluate clinical diagnostic acceptability in hospital settings.
Authors
- Thanyawee Pengpan (ORCID: https://orcid.org/0000-0002-3247-9584)
- Thawatchai Thaikan
- Wirachad Sripoori
- Sowitchaya Huakham
- Raksumon Suksi
Institutions
- Thammasat University (TH)
- King Chulalongkorn Memorial Hospital (TH)
- Naresuan University (TH)
Publication Details
- Journal
- Journal of Applied Clinical Medical Physics
- Published
- 2026-02-01
- DOI
- https://doi.org/10.1002/acm2.70486
- Primary Topic
- Radiation Dose and Imaging
- Type
- article
- Field-Weighted Citation Impact
- 0.00