Monte Carlo simulations of a new 3D electronic detector for radiotherapy quality assurance

<i>Objective</i>. Evaluate the dosimetric performance of a novel three-dimensional (3D) electronic detector array for radiotherapy quality assurance using Monte Carlo simulations.<i>Approach</i>. Monte Carlo simulations were performed with Geant4 and MCNP6.2. A detailed detector model (MC<sub>Model</sub>) was implemented consisting of a 50 × 50 × 50 cm<sup>3</sup>polymethyl methacrylate (PMMA) phantom with 20 imbedded active matrices (AMs) at strategic depths, 1169 pixels per AM, and 23 380 pixels for the entire detector. A pixel comprises a diode, capacitor, and MOSFET, where the diode elements provide a 42 × 42 cm<sup>2</sup>sensitive area within an AM. Photon beam energy spectra of 6 MV and 10 MV, respectively, of a Varian radiotherapy linear accelerator (linac) were used in the simulations. Dosimetric data consisting of per cent depth-doses (PDDs) and cross-/in-plane profiles for field sizes of 5 × 5 cm<sup>2</sup>through 40 × 40 cm<sup>2</sup>were simulated for the MC<sub>Model</sub>, and a homogeneous PMMA phantom (MC<sub>PMMA</sub>) of similar dimensions.<i>Main results.</i>MC<sub>Model</sub>versus MC<sub>PMMA</sub>PDD data difference for Geant4 and MCNP6 were within 3.44% and 3.83%, while profiles (cross-/in-plane) were within 5.54% and 5.68%, for all field sizes and energies.<i>Significance</i>. These results suggest that a 3D electronic detector could provide suitable dosimetric data for radiotherapy QA, and if realised, it could likely provide it in less time than current methods.

Authors

Institutions

Publication Details

Journal
Physics in Medicine and Biology
Published
2026-02-09
DOI
https://doi.org/10.1088/1361-6560/ae43ad
Primary Topic
Advanced Radiotherapy Techniques
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Monte Carlo simulations of a new 3D electronic detector for radiotherapy quality assurance

A. Gonzales Galvez, Mirko Salomón Alva-Sánchez, Alexandre Bonatto, M. Montoya et al.
Physics in Medicine and Biology
Advanced Radiotherapy Techniques
article

Monte Carlo simulations of a new 3D electronic detector for radiotherapy quality assurance

A. Gonzales Galvez, Mirko Salomón Alva-Sánchez, Alexandre Bonatto, M. Montoya, Carmen Sandra Guzmán Calcina, Roger Challco, Enzo Charles Aucca Santiago, Erick Paniagua, Dante Roa, Renzo Eduardo Romero Ataucuri, Miguel Alberto Risco-Castillo, William Souza Santos, Jimmy Hernandez- Bello, Renzo Aaron Ocampo Burga
article en

Abstract

<i>Objective</i>. Evaluate the dosimetric performance of a novel three-dimensional (3D) electronic detector array for radiotherapy quality assurance using Monte Carlo simulations.<i>Approach</i>. Monte Carlo simulations were performed with Geant4 and MCNP6.2. A detailed detector model (MC<sub>Model</sub>) was implemented consisting of a 50 × 50 × 50 cm<sup>3</sup>polymethyl methacrylate (PMMA) phantom with 20 imbedded active matrices (AMs) at strategic depths, 1169 pixels per AM, and 23 380 pixels for the entire detector. A pixel comprises a diode, capacitor, and MOSFET, where the diode elements provide a 42 × 42 cm<sup>2</sup>sensitive area within an AM. Photon beam energy spectra of 6 MV and 10 MV, respectively, of a Varian radiotherapy linear accelerator (linac) were used in the simulations. Dosimetric data consisting of per cent depth-doses (PDDs) and cross-/in-plane profiles for field sizes of 5 × 5 cm<sup>2</sup>through 40 × 40 cm<sup>2</sup>were simulated for the MC<sub>Model</sub>, and a homogeneous PMMA phantom (MC<sub>PMMA</sub>) of similar dimensions.<i>Main results.</i>MC<sub>Model</sub>versus MC<sub>PMMA</sub>PDD data difference for Geant4 and MCNP6 were within 3.44% and 3.83%, while profiles (cross-/in-plane) were within 5.54% and 5.68%, for all field sizes and energies.<i>Significance</i>. These results suggest that a 3D electronic detector could provide suitable dosimetric data for radiotherapy QA, and if realised, it could likely provide it in less time than current methods.

Physics in Medicine and BiologyVol. 71(4)
Federico Villarreal National University (PE), Universidade Federal de Ciências da Saúde de Porto Alegre (BR), Universidade Federal de Sergipe (BR), HealthCare Partners Medical Group (US), Mission Heritage Medical Group (US), Centro Peruano de Estudios Sociales (PE), National University of Engineering (PE), Pontificia Universidad Católica del Perú (PE)
Industry, innovation and infrastructure
Openalex Percentile: Top 55%
Advanced Radiotherapy Techniques
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.