Effect of Crown to Titanium Base Ratio and Force Angle on the Biomechanical Behavior of Dental Implants With CAD/CAM Zirconia and Lithium Disilicate Crowns: A Finite Element Analysis
In anterior single-implant models, the lowest restoration stresses were achieved with a short crown on a 3.5-mm Ti-base and monolithic zirconia. Long crowns (11 mm) increased crown stresses, and abutment height should be tailored to material and anticipated loading direction rather than adjusted by a single rule. Oblique loading consistently raised bone stress compared with 65°, underscoring the need to optimize for axial force transmission.