Optimization of data integration for ultra‐trace zinc isotope ratio analysis by continuous heating thermal ionization mass spectrometry with single‐collector sequential detection

Abstract The thermal ionization mass spectrometry (TIMS) coupled with the continuous heating method was applied to the isotopic analysis of Zn at an ultra‐trace level (1 ng) using the certified reference material IRMM‐3702. Because nanogram‐level Zn produces weak ion signals that require single secondary electron multiplier (SEM) sequential detection, a detailed evaluation of data treatment procedures was carried out to secure high analytical performance. Three integration methods (summed intensities, simple averaging, and intensity‐weighted averaging) were experimentally evaluated for four integration ranges defined relative to the maximum 64 Zn + intensity (full, “over 1%”, “over 25%”, and “over 75%”). The results showed that the analytical performance (accuracy, precision, and uncertainty) is strongly affected by both the integration method and the integration range. Method I (summed intensities) and Method III (intensity‐weighted averaging) produced isotope ratios with relatively high precision and acceptable accuracy, while Method III showed slightly better analytical performance compared to Method I. Method II (simple averaging) produced large deviations and poor precision when applied to the full range, although the analytical performance improved when reduced integration ranges were applied. These findings indicate that the optimized data treatment procedure for the continuous heating TIMS analysis of Zn isotopes is the use of Method I or III with the integration ranges of “over 25%”, as they yield comparable analytical performance in this range. The results establish a methodological procedure for accurate and precise Zn isotope ratio measurements at ultra‐trace levels, extending the applicability of TIMS as a competitive tool for geochemical, environmental, and biomedical research.

Authors

Institutions

Publication Details

Journal
Bulletin of the Korean Chemical Society
Published
2026-02-10
DOI
https://doi.org/10.1002/bkcs.70120
Primary Topic
Isotope Analysis in Ecology
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Optimization of data integration for ultra‐trace zinc isotope ratio analysis by continuous heating thermal ionization mass spectrometry with single‐collector sequential detection

Jong‐Ho Park
Bulletin of the Korean Chemical Society
Isotope Analysis in Ecology
article

Optimization of data integration for ultra‐trace zinc isotope ratio analysis by continuous heating thermal ionization mass spectrometry with single‐collector sequential detection

Jong‐Ho Park
article en

Abstract

Abstract The thermal ionization mass spectrometry (TIMS) coupled with the continuous heating method was applied to the isotopic analysis of Zn at an ultra‐trace level (1 ng) using the certified reference material IRMM‐3702. Because nanogram‐level Zn produces weak ion signals that require single secondary electron multiplier (SEM) sequential detection, a detailed evaluation of data treatment procedures was carried out to secure high analytical performance. Three integration methods (summed intensities, simple averaging, and intensity‐weighted averaging) were experimentally evaluated for four integration ranges defined relative to the maximum 64 Zn + intensity (full, “over 1%”, “over 25%”, and “over 75%”). The results showed that the analytical performance (accuracy, precision, and uncertainty) is strongly affected by both the integration method and the integration range. Method I (summed intensities) and Method III (intensity‐weighted averaging) produced isotope ratios with relatively high precision and acceptable accuracy, while Method III showed slightly better analytical performance compared to Method I. Method II (simple averaging) produced large deviations and poor precision when applied to the full range, although the analytical performance improved when reduced integration ranges were applied. These findings indicate that the optimized data treatment procedure for the continuous heating TIMS analysis of Zn isotopes is the use of Method I or III with the integration ranges of “over 25%”, as they yield comparable analytical performance in this range. The results establish a methodological procedure for accurate and precise Zn isotope ratio measurements at ultra‐trace levels, extending the applicability of TIMS as a competitive tool for geochemical, environmental, and biomedical research.

Bulletin of the Korean Chemical SocietyVol. 47(3)
Jeonju National University of Education (KR), Jeonbuk State Institute (KR), Jeonbuk National University (KR)
Responsible consumption and production
Openalex Percentile: Top 65%
Isotope Analysis in Ecology
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.