Temporal trends and reliability of computed tomography scanner failures: evidence from a real-world time-series analysis

Objective We aimed to characterize temporal and component-specific patterns of CT malfunctions and identifying optimal monitoring intervals are critical to strengthening preventive maintenance and informed equipment management. Methods This study collected failure data from three CT scanners (uCT790, uCT860, uCT960+) at a tertiary hospital, collecting from installation through December 2024. Time-series analysis was used to characterized annual and monthly failure counts, and Pie charts visualized component-level contributions. Reliability performance was quantified using mean time between failures (MTBF), mean time to repair (MTTR), mean time between maintenance (MTBM), mean repair time (MTR), mean maintenance time (MMT), and mean logistics delay time (MLDT). A multi-granularity time-series Poisson-Prophet model (monthly, bi-weekly, weekly) was performed to evaluate predictive accuracy using mean absolute error (MAE). Results From 2019 to 2024, failure trajectories differed across CT scanners. The uCT790 showed a rise to five failures in 2023 followed by a decline to two in 2024. The uCT860 showed a steady increase, reaching six failures in 2024. The uCT960+ increased annually to seven failures in 2024. Failures were concentrated in the scanning table, power supply, and X-ray tube. Reliability analysis showed that the uCT790 achieved the highest MTBF (8554.03 h), the uCT860 the shortest MTTR (6.60 h), and the uCT960+ the lowest MLDT (3.47 h). Across forecasting granularities, predictive accuracy improved with finer granularity: monthly MAE 0.33–1.50, biweekly 0–0.33, and weekly 0–0.17. Conclusion Through integrated reliability assessment and time-series forecasting, this study delineated the distinct failure patterns of three CT scanners and identified bi-weekly forecasting as the most cost-effective temporal resolution, balancing predictive accuracy with resource efficiency.

Authors

Institutions

Publication Details

Journal
Frontiers in Physics
Published
2026-02-10
DOI
https://doi.org/10.3389/fphy.2026.1771484
Primary Topic
Quality and Safety in Healthcare
Type
article
Field-Weighted Citation Impact
0.00

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Temporal trends and reliability of computed tomography scanner failures: evidence from a real-world time-series analysis

Boqi Jia, F. Wang, Rong Li, Ying Xiao et al.
Frontiers in Physics
Quality and Safety in Healthcare
article

Temporal trends and reliability of computed tomography scanner failures: evidence from a real-world time-series analysis

Boqi Jia, F. Wang, Rong Li, Ying Xiao, Feng Xu, Zhenlin Liu, Sujuan Yu, Yun Tian, Shaohua Yin, Xiaoxiao Luan, Ruoyao Pan
article en

Abstract

Objective We aimed to characterize temporal and component-specific patterns of CT malfunctions and identifying optimal monitoring intervals are critical to strengthening preventive maintenance and informed equipment management. Methods This study collected failure data from three CT scanners (uCT790, uCT860, uCT960+) at a tertiary hospital, collecting from installation through December 2024. Time-series analysis was used to characterized annual and monthly failure counts, and Pie charts visualized component-level contributions. Reliability performance was quantified using mean time between failures (MTBF), mean time to repair (MTTR), mean time between maintenance (MTBM), mean repair time (MTR), mean maintenance time (MMT), and mean logistics delay time (MLDT). A multi-granularity time-series Poisson-Prophet model (monthly, bi-weekly, weekly) was performed to evaluate predictive accuracy using mean absolute error (MAE). Results From 2019 to 2024, failure trajectories differed across CT scanners. The uCT790 showed a rise to five failures in 2023 followed by a decline to two in 2024. The uCT860 showed a steady increase, reaching six failures in 2024. The uCT960+ increased annually to seven failures in 2024. Failures were concentrated in the scanning table, power supply, and X-ray tube. Reliability analysis showed that the uCT790 achieved the highest MTBF (8554.03 h), the uCT860 the shortest MTTR (6.60 h), and the uCT960+ the lowest MLDT (3.47 h). Across forecasting granularities, predictive accuracy improved with finer granularity: monthly MAE 0.33–1.50, biweekly 0–0.33, and weekly 0–0.17. Conclusion Through integrated reliability assessment and time-series forecasting, this study delineated the distinct failure patterns of three CT scanners and identified bi-weekly forecasting as the most cost-effective temporal resolution, balancing predictive accuracy with resource efficiency.

Frontiers in PhysicsVol. 14
Peking University (CN), Peking University Third Hospital (CN)
Peking University, Peking University Third Hospital, National Key Research and Development Program of China
Openalex Percentile: Top 68%
Quality and Safety in Healthcare
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.