Navigating preanalytical challenges: a real-world study on single-tube pneumatic tube systems
The single-tube PTS is suitable for transporting blood specimens for biochemical analysis. To maintain the reliability of biochemical results from silica-clot activator tubes, laboratories should minimize pre-transport delays and include resting time as a critical variable in PTS validation protocols.