Retrievability of Fractured Abutment Screws in Dental Implants Using Three Removal Techniques: An In Vitro Pilot Study

<b>Introduction</b>: The fracturing of abutment screws is a recurrent technical complication in implant-supported prostheses that may compromise prosthetic maintenance. Although multiple retrieval approaches have been described, comparative data under controlled experimental conditions remain limited. <b>Materials and Methods</b>: This in vitro pilot study evaluated the retrievability of fractured abutment screws when using three commonly applied instruments: an ultrasonic scaler, a fissure bur, and a screw removal kit. Eighteen implants from a single implant system were embedded in epoxy resin, and abutment screws were fractured under clockwise monotonic torque either with (w/A) or without (w/oA) abutments (<i>n</i>= 3 per retrieval method). Retrieval success and procedure time were recorded. Scanning electron microscopy (SEM) was performed to qualitatively assess deformation of the implant internal hex and screw thread morphology. <b>Results</b>: Fracture torque values were higher in specimens fractured with abutments compared with those without abutments. Fractures induced without abutments appeared to extend deeper within the screw channel, engaging a greater number of internal threads. In this pilot study, a shorter retrieval time was observed with the screw removal kit and fissure bur compared with the ultrasonic scaler, although retrieval outcomes varied between specimens. SEM observations suggested differing patterns of internal hex deformation between the retrieval techniques. <b>Conclusions</b>: Within the limitations of this in vitro pilot study, different retrieval approaches demonstrated characteristic mechanical behaviors and deformation patterns in the implant internal connection. These preliminary findings provide descriptive insight into the retrievability of fractured screws and may serve as a basis for future studies with larger sample sizes and clinically relevant fracture models.

Authors

Institutions

Publication Details

Journal
Journal of Functional Biomaterials
Published
2026-02-09
DOI
https://doi.org/10.3390/jfb17020085
Primary Topic
Dental Implant Techniques and Outcomes
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Retrievability of Fractured Abutment Screws in Dental Implants Using Three Removal Techniques: An In Vitro Pilot Study

Ming-Dih Jeng, Tzu-Yun Huang, Amber Yeh Jeng
Journal of Functional Biomaterials
Dental Implant Techniques and Outcomes
article

Retrievability of Fractured Abutment Screws in Dental Implants Using Three Removal Techniques: An In Vitro Pilot Study

Ming-Dih Jeng, Tzu-Yun Huang, Amber Yeh Jeng
article en

Abstract

<b>Introduction</b>: The fracturing of abutment screws is a recurrent technical complication in implant-supported prostheses that may compromise prosthetic maintenance. Although multiple retrieval approaches have been described, comparative data under controlled experimental conditions remain limited. <b>Materials and Methods</b>: This in vitro pilot study evaluated the retrievability of fractured abutment screws when using three commonly applied instruments: an ultrasonic scaler, a fissure bur, and a screw removal kit. Eighteen implants from a single implant system were embedded in epoxy resin, and abutment screws were fractured under clockwise monotonic torque either with (w/A) or without (w/oA) abutments (<i>n</i>= 3 per retrieval method). Retrieval success and procedure time were recorded. Scanning electron microscopy (SEM) was performed to qualitatively assess deformation of the implant internal hex and screw thread morphology. <b>Results</b>: Fracture torque values were higher in specimens fractured with abutments compared with those without abutments. Fractures induced without abutments appeared to extend deeper within the screw channel, engaging a greater number of internal threads. In this pilot study, a shorter retrieval time was observed with the screw removal kit and fissure bur compared with the ultrasonic scaler, although retrieval outcomes varied between specimens. SEM observations suggested differing patterns of internal hex deformation between the retrieval techniques. <b>Conclusions</b>: Within the limitations of this in vitro pilot study, different retrieval approaches demonstrated characteristic mechanical behaviors and deformation patterns in the implant internal connection. These preliminary findings provide descriptive insight into the retrievability of fractured screws and may serve as a basis for future studies with larger sample sizes and clinically relevant fracture models.

Journal of Functional BiomaterialsVol. 17(2)
Kaohsiung Medical University (TW), Far Eastern Memorial Hospital (TW)
Openalex Percentile: Top 62%
Dental Implant Techniques and Outcomes
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.