Cardiac MR function analysis with DL-based super resolution reconstruction: application in the clinical setting

To assess differences in volumetry, image quality and acquisition time between balanced steady-state free precession cine sequences acquired using (a) a standardized sensitivity encoding (SENSE) approach and (b) deep learning-based super-resolution reconstruction based on high-resolution images acquired with compressed sensitivity encoding (C-SENSE). We retrospectively evaluated 31 consecutive patients (mean age 61.2 ± 13.1 years, 26% female (8/31) and 74% male (23/31)) undergoing cardiac magnetic resonance imaging (MRI) examinations to assess for the presence of ischemic and non-ischemic cardiomyopathies. Cine images were acquired using a 1.5T Philips Ingenia MRI scanner, with classic parallel imaging (SENSE) and compressed sensing (C-SENSE) accelerated acquisition techniques (R = 2 and R = 4, respectively). C-SENSE datasets were reconstructed using a deep learning-based denoising and super-resolution algorithm to enhance image resolution and quality (CS-SR). To evaluate cardiac function, manual left ventricular (LV) segmentation and volumetric analysis were performed on both datasets by two readers, who were blinded to the clinical data. Image quality was rated independently by three readers using Likert scales. Correlation between SENSE and CS-SR datasets with respect to LV volumetry was high (r = 0.98-1.00), with no significant differences found for end-diastolic volume (mean difference 0.04 ml, limits of agreement (LoA) -11.19 to 11.26 ml; p = 0.970) or end-systolic volume (mean difference 1.60 ml, LoA - 7.48 to 10.68 ml; p = 0.064). Overall subjective image quality was comparable (p = 0.061), with CS-SR offering better image sharpness at the cost of increased artifacts (p < 0.001 respectively). Image acquisition time was significantly accelerated with C-SENSE acquisition (SENSE: 411.1 ± 47.7 s, C-SENSE: 165.6 ± 21.5 s; p < 0.001). CS-SR shows promise in streamlining routine cardiac imaging by significantly shortening acquisition times, without impairing LV volumetric analysis, while preserving overall image quality and resolution.

Authors

Institutions

Publication Details

Journal
The International Journal of Cardiovascular Imaging
Published
2026-02-09
DOI
https://doi.org/10.1007/s10554-026-03642-8
Primary Topic
Cardiac Imaging and Diagnostics
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Cardiac MR function analysis with DL-based super resolution reconstruction: application in the clinical setting

Matthias Eberhard, Franziska Adomat, Lucas Weber, Tobias Hoh et al.
The International Journal of Cardiovascular Imaging
Cardiac Imaging and Diagnostics
article

Cardiac MR function analysis with DL-based super resolution reconstruction: application in the clinical setting

Matthias Eberhard, Franziska Adomat, Lucas Weber, Tobias Hoh, R. H. Manka, Christof Schaub, Xenia Fischer
article en

Abstract

To assess differences in volumetry, image quality and acquisition time between balanced steady-state free precession cine sequences acquired using (a) a standardized sensitivity encoding (SENSE) approach and (b) deep learning-based super-resolution reconstruction based on high-resolution images acquired with compressed sensitivity encoding (C-SENSE). We retrospectively evaluated 31 consecutive patients (mean age 61.2 ± 13.1 years, 26% female (8/31) and 74% male (23/31)) undergoing cardiac magnetic resonance imaging (MRI) examinations to assess for the presence of ischemic and non-ischemic cardiomyopathies. Cine images were acquired using a 1.5T Philips Ingenia MRI scanner, with classic parallel imaging (SENSE) and compressed sensing (C-SENSE) accelerated acquisition techniques (R = 2 and R = 4, respectively). C-SENSE datasets were reconstructed using a deep learning-based denoising and super-resolution algorithm to enhance image resolution and quality (CS-SR). To evaluate cardiac function, manual left ventricular (LV) segmentation and volumetric analysis were performed on both datasets by two readers, who were blinded to the clinical data. Image quality was rated independently by three readers using Likert scales. Correlation between SENSE and CS-SR datasets with respect to LV volumetry was high (r = 0.98-1.00), with no significant differences found for end-diastolic volume (mean difference 0.04 ml, limits of agreement (LoA) -11.19 to 11.26 ml; p = 0.970) or end-systolic volume (mean difference 1.60 ml, LoA - 7.48 to 10.68 ml; p = 0.064). Overall subjective image quality was comparable (p = 0.061), with CS-SR offering better image sharpness at the cost of increased artifacts (p < 0.001 respectively). Image acquisition time was significantly accelerated with C-SENSE acquisition (SENSE: 411.1 ± 47.7 s, C-SENSE: 165.6 ± 21.5 s; p < 0.001). CS-SR shows promise in streamlining routine cardiac imaging by significantly shortening acquisition times, without impairing LV volumetric analysis, while preserving overall image quality and resolution.

The International Journal of Cardiovascular Imaging
Philips (Finland) (FI), University Hospital of Bern (CH), Kantonsspital Winterthur (CH), University Hospital of Zurich (CH), Department of Health Canton of Zurich (CH)
Openalex Percentile: Top 63%
Cardiac Imaging and Diagnostics
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.