Dosimetric characterization of scatter foil‐enhanced contact collimation for small superficial electron beam therapy
Thin metallic foils placed on Cerrobend skin collimators enable a controllable balance between dose uniformity (improved with Al) and conformality with distal sparing (enhanced with Pb). This simple, LINAC-independent configuration offers a cost-effective method for modulating small-field electron beam characteristics and may serve as a practical adjunct for treating superficial lesions.
Authors
- Abdulaziz Alhussan (ORCID: https://orcid.org/0000-0002-9920-2803)
- Richard J. Crilly
Institutions
- Oregon Health & Science University (US)
Publication Details
- Journal
- Journal of Applied Clinical Medical Physics
- Published
- 2026-02-01
- DOI
- https://doi.org/10.1002/acm2.70484
- Primary Topic
- Advanced Radiotherapy Techniques
- Type
- article
- Field-Weighted Citation Impact
- 0.00