Performance of two portable exhaled nitric oxide fraction devices compared to a "gold standard" chemiluminescence device

Background: Recent guidelines for the diagnosis of asthma in schoolchildren recommend the measurement of exhaled nitric oxide fraction (F$_{ENO}$) as part of the diagnostic algorithm. However, implementation may be hampered by the lack of F$_{ENO}$ devices that are affordable and usable in all healthcare settings. We aimed to compare the performance of two portable F$_{ENO}$ devices (Evernoa (EVE) and NObreath (NOB)) to a stationary "gold standard" device (CLD 88 sp (CLD)) in children. Methods: 106 children aged 6-17 years under investigation or monitoring for asthma underwent F$_{ENO}$ measurements using the three devices in randomised order. Results: All devices showed high repeatability across a wide F$_{ENO}$ range (2.5-191.9 ppb). Median (interquartile range) F ENO levels were significantly lower with the portable devices (20.8 (9.5-43.5), 16 (9-36) and 22.8 (13.2-55.2) ppb for NOB, EVE and CLD, respectively; p<0.0001). Despite the proportional bias (-20% (NOB) and -40% (EVE)), both portable devices demonstrated good overall agreement with CLD (>94%) at a cut-off level of 25 ppb but lower agreement for a cut-off of 35 ppb. EVE required a greater number of attempts compared to NOB and CLD to achieve two valid measurements. Conclusions: Both portable devices showed limited interchangeability with the gold standard, making them less applicable for research and disease monitoring purposes. However, good overall agreement at the European Respiratory Society cut-off level (25 ppb) suggests potential as a simple and convenient screening tool in clinical settings for initial asthma diagnosis.

Authors

Publication Details

Journal
Universität Zürich, ZORA
Published
2025-04-28
DOI
https://doi.org/10.5167/uzh-291012
Primary Topic
Asthma and respiratory diseases
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Performance of two portable exhaled nitric oxide fraction devices compared to a "gold standard" chemiluminescence device

Srdjan Micic, Leo Thanikkel, Simon Vogt, Alexander Möller et al.
Universität Zürich, ZORA
Asthma and respiratory diseases
article

Performance of two portable exhaled nitric oxide fraction devices compared to a "gold standard" chemiluminescence device

Srdjan Micic, Leo Thanikkel, Simon Vogt, Alexander Möller, Nadja Delliehausen, Andreas Vogt, Zurich Pediatric Exhalomics Group
article en

Abstract

Background: Recent guidelines for the diagnosis of asthma in schoolchildren recommend the measurement of exhaled nitric oxide fraction (F$_{ENO}$) as part of the diagnostic algorithm. However, implementation may be hampered by the lack of F$_{ENO}$ devices that are affordable and usable in all healthcare settings. We aimed to compare the performance of two portable F$_{ENO}$ devices (Evernoa (EVE) and NObreath (NOB)) to a stationary "gold standard" device (CLD 88 sp (CLD)) in children. Methods: 106 children aged 6-17 years under investigation or monitoring for asthma underwent F$_{ENO}$ measurements using the three devices in randomised order. Results: All devices showed high repeatability across a wide F$_{ENO}$ range (2.5-191.9 ppb). Median (interquartile range) F ENO levels were significantly lower with the portable devices (20.8 (9.5-43.5), 16 (9-36) and 22.8 (13.2-55.2) ppb for NOB, EVE and CLD, respectively; p<0.0001). Despite the proportional bias (-20% (NOB) and -40% (EVE)), both portable devices demonstrated good overall agreement with CLD (>94%) at a cut-off level of 25 ppb but lower agreement for a cut-off of 35 ppb. EVE required a greater number of attempts compared to NOB and CLD to achieve two valid measurements. Conclusions: Both portable devices showed limited interchangeability with the gold standard, making them less applicable for research and disease monitoring purposes. However, good overall agreement at the European Respiratory Society cut-off level (25 ppb) suggests potential as a simple and convenient screening tool in clinical settings for initial asthma diagnosis.

Universität Zürich, ZORA
Openalex Percentile: Top 28%
Asthma and respiratory diseases
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.