Quantification of Myocardial Iron and Fat — An Experimental Study with Photon-Counting Detector CT

Objectives: To determine the feasibility and accuracy of photon-counting detector (PCD)-CT for iron and fat quantification in the myocardium. Materials and Methods: Cylindrical tubes were filled with porcine myocardium and iron-citrate with iron concentrations of 0-20mg Fe/g. Dilution series were prepared with myocardium and iron (no-fat probes) and with 5%-fat (fat probes). The tubes were positioned in a chest phantom and were scanned with a calcium-scoring protocol on a PCD-CT. A re-parameterized three-material decomposition was used to separate iron and fat from myocardium. Results: On virtual monoenergetic images, attenuation increased linearly with iron concentrations in both fat and no-fat probes. In no-fat probes, linear regression yielded a slope of 1.2HU/(mgFe/g) with an intercept of 35.8HU (R2=0.964). In the fat probes, the slope was similar at 1.1HU/(mgFe/g), while the regression line shifted downwards by 6.1HU with an intercept of 29.6HU (R2=0.985). Iron maps separated fat from iron with calculated median fat-fractions of 4.85 in the fat and 0.90 in the no-fat probes. In iron images, attenuation increased linearly with increasing iron concentrations, with similar slopes between fat and no-fat probes and negligible differences in the intercept. Conclusion: Experimental evidence indicates the feasibility and accuracy of PCD-CT for iron and fat quantification in the myocardium. Iron-specific three-material decomposition eliminates the confounding effect of fat on myocardial iron quantification. Advances in Knowledge This study highlights the value of dual-energy CT with three-material decomposition for quantifying iron and fat in the myocardium. Thus, CT could serve as alternative for the current reference standard MRI.

Authors

Publication Details

Journal
Universität Zürich, ZORA
Published
2026-01-25
DOI
https://doi.org/10.5167/uzh-291009
Primary Topic
Advanced X-ray and CT Imaging
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Quantification of Myocardial Iron and Fat — An Experimental Study with Photon-Counting Detector CT

Johannes Froehlich, Matthias Eberhard, Konstantin Klambauer, Philipp Maintz et al.
Universität Zürich, ZORA
Advanced X-ray and CT Imaging
article

Quantification of Myocardial Iron and Fat — An Experimental Study with Photon-Counting Detector CT

Johannes Froehlich, Matthias Eberhard, Konstantin Klambauer, Philipp Maintz, Tristan T. Demmert, Hatem Alkadhi, Victor Mergen, Lukas J Moser, Thomas Flohr
article en

Abstract

Objectives: To determine the feasibility and accuracy of photon-counting detector (PCD)-CT for iron and fat quantification in the myocardium. Materials and Methods: Cylindrical tubes were filled with porcine myocardium and iron-citrate with iron concentrations of 0-20mg Fe/g. Dilution series were prepared with myocardium and iron (no-fat probes) and with 5%-fat (fat probes). The tubes were positioned in a chest phantom and were scanned with a calcium-scoring protocol on a PCD-CT. A re-parameterized three-material decomposition was used to separate iron and fat from myocardium. Results: On virtual monoenergetic images, attenuation increased linearly with iron concentrations in both fat and no-fat probes. In no-fat probes, linear regression yielded a slope of 1.2HU/(mgFe/g) with an intercept of 35.8HU (R2=0.964). In the fat probes, the slope was similar at 1.1HU/(mgFe/g), while the regression line shifted downwards by 6.1HU with an intercept of 29.6HU (R2=0.985). Iron maps separated fat from iron with calculated median fat-fractions of 4.85 in the fat and 0.90 in the no-fat probes. In iron images, attenuation increased linearly with increasing iron concentrations, with similar slopes between fat and no-fat probes and negligible differences in the intercept. Conclusion: Experimental evidence indicates the feasibility and accuracy of PCD-CT for iron and fat quantification in the myocardium. Iron-specific three-material decomposition eliminates the confounding effect of fat on myocardial iron quantification. Advances in Knowledge This study highlights the value of dual-energy CT with three-material decomposition for quantifying iron and fat in the myocardium. Thus, CT could serve as alternative for the current reference standard MRI.

Universität Zürich, ZORA
Openalex Percentile: Top 65%
Advanced X-ray and CT Imaging
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.