Improved electron paramagnetic resonance spectroscopy sensitivity for aqueous biological samples using low-volume multi-channel cells and dielectric resonators

Reducing sample volumes for electron paramagnetic resonance (EPR) spectroscopy applications places increasing demands on hardware design to preserve or enhance EPR signal intensity. This work presents the design, fabrication, and testing of dielectric resonators and multi-channel aqueous sample cells for applications in X-band (nominally 9.5 GHz) EPR. Our aim was to maximize the EPR signal intensity for sample sizes of 3-4 μl and 200 nl. These advances are summarized as follows: single-crystal sapphire and rutile dielectric resonators with very low loss tangent and high resonator efficiency; minimum dielectric resonator coupling to radiation shield to reduce ohmic losses; 3D-printed aqueous sample cells with thin multi-channel construction to minimize radio frequency dissipation in the sample; and a Gordon coupler for maximum coupling range and minimum stored energy to eliminate frequency shifts during tuning. Sample tube cross sections were designed by leveraging insights gained from analytic theory to inform finite-element modeling of electromagnetic fields. Experimental comparisons of multi-channel sample cells using a sapphire resonator exhibited a 2.2-fold increase in EPR signal intensity compared with a standard capillary at 3-4 μl, while simulations predict an additional 23% improvement with further 3D printing advances. For samples at 200 nl, a rutile dielectric resonator with a multi-channel sample cell was simulated to improve EPR sensitivity by a 2.7-fold increase compared with a capillary at the same volume.

Authors

Institutions

Publication Details

Journal
Review of Scientific Instruments
Published
2026-02-01
DOI
https://doi.org/10.1063/5.0285590
Primary Topic
Electron Spin Resonance Studies
Type
article
Field-Weighted Citation Impact
0.00

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Improved electron paramagnetic resonance spectroscopy sensitivity for aqueous biological samples using low-volume multi-channel cells and dielectric resonators

Candice S. Klug, Richard R. Mett, Michael T. Lerch, Jason W. Sidabras et al.
Review of Scientific Instruments
Electron Spin Resonance Studies
article

Improved electron paramagnetic resonance spectroscopy sensitivity for aqueous biological samples using low-volume multi-channel cells and dielectric resonators

Candice S. Klug, Richard R. Mett, Michael T. Lerch, Jason W. Sidabras, Alexander M. Garces, Joseph T. Wehrley, Anand Anilkumar
article en

Abstract

Reducing sample volumes for electron paramagnetic resonance (EPR) spectroscopy applications places increasing demands on hardware design to preserve or enhance EPR signal intensity. This work presents the design, fabrication, and testing of dielectric resonators and multi-channel aqueous sample cells for applications in X-band (nominally 9.5 GHz) EPR. Our aim was to maximize the EPR signal intensity for sample sizes of 3-4 μl and 200 nl. These advances are summarized as follows: single-crystal sapphire and rutile dielectric resonators with very low loss tangent and high resonator efficiency; minimum dielectric resonator coupling to radiation shield to reduce ohmic losses; 3D-printed aqueous sample cells with thin multi-channel construction to minimize radio frequency dissipation in the sample; and a Gordon coupler for maximum coupling range and minimum stored energy to eliminate frequency shifts during tuning. Sample tube cross sections were designed by leveraging insights gained from analytic theory to inform finite-element modeling of electromagnetic fields. Experimental comparisons of multi-channel sample cells using a sapphire resonator exhibited a 2.2-fold increase in EPR signal intensity compared with a standard capillary at 3-4 μl, while simulations predict an additional 23% improvement with further 3D printing advances. For samples at 200 nl, a rutile dielectric resonator with a multi-channel sample cell was simulated to improve EPR sensitivity by a 2.7-fold increase compared with a capillary at the same volume.

Review of Scientific InstrumentsVol. 97(2)
Medical College of Wisconsin (US), Milwaukee School of Engineering (US)
National Institutes of Health, National Institute of General Medical Sciences
Affordable and clean energy
Openalex Percentile: Top 68%
Electron Spin Resonance Studies
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.