Pharmacokinetic Modeling of Ticagrelor and Ticagrelor Active Metabolite Removal by the DrugSorb-ATR Device in a Randomized Controlled Study of Patients Undergoing Cardiac Surgery

Excessive surgical bleeding is a potential risk in patients taking ticagrelor who must undergo urgent cardiothoracic (CT) surgery before adequate washout of the drug can occur. The DrugSorb-Antithrombotic removal (ATR) device is a polymer sorbent-filled hemoadsorption cartridge that can remove unbound (active) fractions of ticagrelor and ticagrelor active metabolite (TAM) from blood. STAR-T was a randomized double-blind sham-controlled clinical trial investigating whether the intraoperative use of the device could reduce perioperative bleeding complications in patients undergoing CT surgery within 2 days of ticagrelor discontinuation. Blood samples were collected during the study for total drug level measurements because the ability to measure unbound ticagrelor and TAM (0.2% of total levels) requires an ultra-high sensitivity assay, which is not commercially available. A published and validated pharmacokinetic (PK) model was used to explore the effect of the device on unbound ticagrelor/TAM using the total drug concentrations from the study. The model performed well for simulations of total ticagrelor and TAM, which indicated that the unbound concentrations were also appropriate. The model demonstrated that DrugSorb-ATR significantly reduced unbound ticagrelor and TAM concentrations. Linear and logistic regression analyses of summed ticagrelor and TAM concentrations showed that the DrugSorb-ATR device reduced the probability of clinically relevant bleeding in STAR-T because of the reduction in unbound ticagrelor and TAM.

Authors

Institutions

Publication Details

Journal
Journal of Cardiovascular Pharmacology
Published
2026-02-01
DOI
https://doi.org/10.1097/fjc.0000000000001779
Primary Topic
Antiplatelet Therapy and Cardiovascular Diseases
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Pharmacokinetic Modeling of Ticagrelor and Ticagrelor Active Metabolite Removal by the DrugSorb-ATR Device in a Randomized Controlled Study of Patients Undergoing Cardiac Surgery

Efthymios Deliargyris, Diane R. Mould, David J. Schneider, Kevin Sweeney et al.
Journal of Cardiovascular Pharmacology
Antiplatelet Therapy and Cardiovascular Diseases
article

Pharmacokinetic Modeling of Ticagrelor and Ticagrelor Active Metabolite Removal by the DrugSorb-ATR Device in a Randomized Controlled Study of Patients Undergoing Cardiac Surgery

Efthymios Deliargyris, Diane R. Mould, David J. Schneider, Kevin Sweeney, Jayne Prats, Weihong Fan
article en

Abstract

Excessive surgical bleeding is a potential risk in patients taking ticagrelor who must undergo urgent cardiothoracic (CT) surgery before adequate washout of the drug can occur. The DrugSorb-Antithrombotic removal (ATR) device is a polymer sorbent-filled hemoadsorption cartridge that can remove unbound (active) fractions of ticagrelor and ticagrelor active metabolite (TAM) from blood. STAR-T was a randomized double-blind sham-controlled clinical trial investigating whether the intraoperative use of the device could reduce perioperative bleeding complications in patients undergoing CT surgery within 2 days of ticagrelor discontinuation. Blood samples were collected during the study for total drug level measurements because the ability to measure unbound ticagrelor and TAM (0.2% of total levels) requires an ultra-high sensitivity assay, which is not commercially available. A published and validated pharmacokinetic (PK) model was used to explore the effect of the device on unbound ticagrelor/TAM using the total drug concentrations from the study. The model performed well for simulations of total ticagrelor and TAM, which indicated that the unbound concentrations were also appropriate. The model demonstrated that DrugSorb-ATR significantly reduced unbound ticagrelor and TAM concentrations. Linear and logistic regression analyses of summed ticagrelor and TAM concentrations showed that the DrugSorb-ATR device reduced the probability of clinically relevant bleeding in STAR-T because of the reduction in unbound ticagrelor and TAM.

Journal of Cardiovascular PharmacologyVol. 87(2)
University of Vermont (US), Lynn University (US), CytoSorbents (United States) (US), Phoenixville Hospital (US), University of Sheffield (GB), Unisys (United States) (US)
Good health and well-being
Openalex Percentile: Top 66%
Antiplatelet Therapy and Cardiovascular Diseases
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.