Injection method alters diagnostic confidence and quantitative metrics in [18F]fluorocholine PET/CT for parathyroid adenoma localization

Studies evaluating [18F]fluorocholine PET/CT for parathyroid adenoma localisation fail to report the injection method used. At our department, manual injection method (MIM) was initially used, before converting to an automated injection system (AIS). Afterwards, physicians observed increased retention of the radiotracer within the veins leading from the injection site, which had not been as present previously. The aim of this study was to evaluate whether the injection method had a clinical impact. The study compared two groups consisting of 60 patients, where the injection method was the only difference. Two physicians, blinded to injection method, evaluated if retention was present and interpretation was jeopardised. Quantitative measures were determined based on the original diagnostic confidence scores (DCS) for each suspected lesion, as these were compared between patient groups. Additionally, the maximum standardised uptake values (SUVmax) were measured for the suspected lesions along with the thyroid and salivary glands as a physiological reference point. Comparisons of the quantitative parameters were also performed between patients with or without retention. There was significant more occurrence of retention in AIS-group (83.3%) than MIM group (36.7%), (p < 0.00001). Retention did not influence the interpretation of any patient. A significant difference of DCS was observed, as 18% of lesions in the AIS-group had a low score, compared to only 1% in the MIM-group (p < 0.001). SUVmax was significantly lower in the AIS-group than the MIM-group for the reported adenomas (10.1 vs. 11.8, p = 0.008). In addition, SUVmax in the AIS-group was significantly lower in both the thyroid (4.6 vs. 5.2, p = 0.023) and salivary glands (10.8 vs. 12.4, p < 0.001). Comparison between patients with and without retention revealed similar differences in DCS and SUVmax for adenomas. The AIS is associated with a significant higher presence of retention, low diagnostic confidence and overall lower SUVmax, which indicates an aspect of clinical impact. It could potentially lead to a parathyroid adenoma being missed or classified as normal parathyroid tissue. Thus, manual injection method might be the method of choice, to ensure the most optimal clinical outcome.

Authors

Institutions

Publication Details

Journal
EJNMMI Reports
Published
2026-02-05
DOI
https://doi.org/10.1186/s41824-026-00285-9
Primary Topic
Parathyroid Disorders and Treatments
Type
article
Field-Weighted Citation Impact
0.00

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Injection method alters diagnostic confidence and quantitative metrics in [18F]fluorocholine PET/CT for parathyroid adenoma localization

Henriette Juel Nissen, Peymaneh Abadi, Hanne Reilev Møller, Mette Rode et al.
EJNMMI Reports
Parathyroid Disorders and Treatments
article

Injection method alters diagnostic confidence and quantitative metrics in [18F]fluorocholine PET/CT for parathyroid adenoma localization

Henriette Juel Nissen, Peymaneh Abadi, Hanne Reilev Møller, Mette Rode, Paw Christian Holdgaard
article en

Abstract

Studies evaluating [18F]fluorocholine PET/CT for parathyroid adenoma localisation fail to report the injection method used. At our department, manual injection method (MIM) was initially used, before converting to an automated injection system (AIS). Afterwards, physicians observed increased retention of the radiotracer within the veins leading from the injection site, which had not been as present previously. The aim of this study was to evaluate whether the injection method had a clinical impact. The study compared two groups consisting of 60 patients, where the injection method was the only difference. Two physicians, blinded to injection method, evaluated if retention was present and interpretation was jeopardised. Quantitative measures were determined based on the original diagnostic confidence scores (DCS) for each suspected lesion, as these were compared between patient groups. Additionally, the maximum standardised uptake values (SUVmax) were measured for the suspected lesions along with the thyroid and salivary glands as a physiological reference point. Comparisons of the quantitative parameters were also performed between patients with or without retention. There was significant more occurrence of retention in AIS-group (83.3%) than MIM group (36.7%), (p < 0.00001). Retention did not influence the interpretation of any patient. A significant difference of DCS was observed, as 18% of lesions in the AIS-group had a low score, compared to only 1% in the MIM-group (p < 0.001). SUVmax was significantly lower in the AIS-group than the MIM-group for the reported adenomas (10.1 vs. 11.8, p = 0.008). In addition, SUVmax in the AIS-group was significantly lower in both the thyroid (4.6 vs. 5.2, p = 0.023) and salivary glands (10.8 vs. 12.4, p < 0.001). Comparison between patients with and without retention revealed similar differences in DCS and SUVmax for adenomas. The AIS is associated with a significant higher presence of retention, low diagnostic confidence and overall lower SUVmax, which indicates an aspect of clinical impact. It could potentially lead to a parathyroid adenoma being missed or classified as normal parathyroid tissue. Thus, manual injection method might be the method of choice, to ensure the most optimal clinical outcome.

EJNMMI ReportsVol. 10(1)
Lillebaelt Hospital (DK)
Odense Universitetshospital, Syddansk Universitet
Good health and well-being
Openalex Percentile: Top 69%
Parathyroid Disorders and Treatments
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.