Comparative in vitro evaluation of the mechanical behavior of implant–abutment connections for zirconia and titanium implants

Dental implants have significantly improved the rehabilitation of edentulous spaces. With the growing demand for metal-free treatments, zirconia has emerged as an alternative to titanium. While titanium implant connections are well documented, limited evidence exists regarding the mechanical behavior and prosthetic screw loosening of zirconia implants. The Neodent Zi® implant with the ZiLock® ceramic connection remains poorly investigated, particularly in comparison with conventional titanium systems. This study aimed to compare the removal torque values between prosthetic screws in titanium morse taper connection and zirconia ZiLock® connection implants. Twenty implants were divided into two groups: 10 Morse taper connection titanium implants (Neodent® Alvim®) and 10 ZiLock® connection zirconia implants (Neodent® Zi®). Both groups underwent initial torquing, mechanical cycling following ISO 14801 standards, and subsequent removal torque value measurement. Structural analysis of the implants was conducted using a stereomicroscope. Data was presented using descriptive statistics including means, standard deviations (SD), and frequencies. The Shapiro-Wilk test was performed to assess data normality and the Student’s t-test was applied to evaluate differences between groups. Categorical data were subjected to Fisher’s exact test. Statistical significance was established as p < 0.05. The titanium group showed higher mean removal torque values (38.27 Ncm) compared to the zirconia group (25.65 Ncm), with statistically significant differences (p < 0.001). The zirconia group exhibited fractures in two abutments. Zirconia implants with the ZiLock® connection displayed lower removal torque values than titanium implants but within acceptable clinical limits.

Authors

Institutions

Publication Details

Journal
BMC Oral Health
Published
2026-02-05
DOI
https://doi.org/10.1186/s12903-026-07855-0
Primary Topic
Dental Implant Techniques and Outcomes
Type
article
Field-Weighted Citation Impact
0.00

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Comparative in vitro evaluation of the mechanical behavior of implant–abutment connections for zirconia and titanium implants

Bárbara Magalhães Figueiredo Dias, Camila da Silva Resende, Lara Rúbia Marques Nascimento, Frederico Santos Lages et al.
BMC Oral Health
Dental Implant Techniques and Outcomes
article

Comparative in vitro evaluation of the mechanical behavior of implant–abutment connections for zirconia and titanium implants

Bárbara Magalhães Figueiredo Dias, Camila da Silva Resende, Lara Rúbia Marques Nascimento, Frederico Santos Lages, Dhelfeson Willya Douglas-de-Oliveira, Rafaela Regina de Lima, Olívia Francescato
article en

Abstract

Dental implants have significantly improved the rehabilitation of edentulous spaces. With the growing demand for metal-free treatments, zirconia has emerged as an alternative to titanium. While titanium implant connections are well documented, limited evidence exists regarding the mechanical behavior and prosthetic screw loosening of zirconia implants. The Neodent Zi® implant with the ZiLock® ceramic connection remains poorly investigated, particularly in comparison with conventional titanium systems. This study aimed to compare the removal torque values between prosthetic screws in titanium morse taper connection and zirconia ZiLock® connection implants. Twenty implants were divided into two groups: 10 Morse taper connection titanium implants (Neodent® Alvim®) and 10 ZiLock® connection zirconia implants (Neodent® Zi®). Both groups underwent initial torquing, mechanical cycling following ISO 14801 standards, and subsequent removal torque value measurement. Structural analysis of the implants was conducted using a stereomicroscope. Data was presented using descriptive statistics including means, standard deviations (SD), and frequencies. The Shapiro-Wilk test was performed to assess data normality and the Student’s t-test was applied to evaluate differences between groups. Categorical data were subjected to Fisher’s exact test. Statistical significance was established as p < 0.05. The titanium group showed higher mean removal torque values (38.27 Ncm) compared to the zirconia group (25.65 Ncm), with statistically significant differences (p < 0.001). The zirconia group exhibited fractures in two abutments. Zirconia implants with the ZiLock® connection displayed lower removal torque values than titanium implants but within acceptable clinical limits.

BMC Oral HealthVol. 26(1)
Universidade Federal de Minas Gerais (BR), Universidade Federal dos Vales do Jequitinhonha e Mucuri (BR)
Coordenação de Aperfeiçoamento de Pessoal de Nível Superior, Fundação de Amparo à Pesquisa do Estado de Minas Gerais
Openalex Percentile: Top 66%
Dental Implant Techniques and Outcomes
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.