Diagnostic performance evaluation of the Abbott Bioline™ HIV/Syphilis Duo rapid test at the national public health laboratory, Yaounde-Cameroon

The Bioline™ HIV/Syphilis Duo rapid diagnostic test (RDT) is the first World Health Organization (WHO) prequalified Duo RDT for the simultaneous detection of human immunodeficiency virus (HIV) and Treponema pallidum antibodies in human blood samples. Several studies have shown the satisfactory diagnostic performance of this test among vulnerable populations such as pregnant women. However, the diagnostic performance of this Duo test is scarce in the general population in Cameroon. The present study aimed to assess the diagnostic performance of the Abbot Bioline™ HIV/Syphilis Duo RDT in Cameroon. A laboratory based cross-sectional and comparative study was conducted from January, 6th to June, 19th 2024 on serum/plasma collected from adult individuals (aged: ≥ 21 years), recruited consecutively in health facilities in eight regions of Cameroon. The blood samples were received and stored at the National Public Health Laboratory (NPHL). The Bioline™ HIV 1/2 test was performed following the national testing algorithm with Determine™ HIV 1/2 (first test) and KHB Shanghai HIV1/2 (second test). Meanwhile, the performance of the Bioline™ syphilis test was compared with the Hightop RDT and Treponema pallidum hemagglutination assay (TPHA). The enzyme linked immunosorbent assay (ELISA) was used as the reference assay. Sensitivity, specificity, positive predictive value (PPV), negative predictive value (NPV), accuracy and Cohen’s kappa (κ) value were calculated for each marker against the reference assay within a 95% confidence interval (CI) using Epi-info version 7, MedCalc and QuickCalcs for online statistical analysis. A total of 534 blood samples were tested at the NPHL with 70.97% (379/534) from females and 29.03% (155/534) from males. For HIV diagnosis, the Bioline™ HIV1/2 test had a sensitivity of 99.70% (95% CI: 98.30–99.9), specificity of 98.5% (95% CI: 95.58–99.5), PPV of 99.1% (95% CI: 97.40–99.7), NPV of 99.5% (95% CI: 97.3–99.9), accuracy of 99.1%(95% CI: 97-99.7) and Cohen’s kappa of 98.4%(95% CI:96.7–99.9). For syphilis diagnosis, the Bioline™ test had a sensitivity of 95% (95% CI: 73.7–99.7), specificity of 100% (95% CI: 99.1–100), PPV of 100% (95% CI: 79.1–100), NPV of 99.8% (95% CI: 98.7–100), accuracy of 99.9% (95% CI: 99–100), and Cohen’s kappa of 97.3%(95% CI:92.1–100). The results of this study show that the Abbot Bioline™ HIV/Syphilis Duo RDT has a good performance. Therefore, we suggest that this test could be suitable for use in the diagnosis of HIV and syphilis.

Authors

Institutions

Publication Details

Journal
BMC Infectious Diseases
Published
2026-02-04
DOI
https://doi.org/10.1186/s12879-026-12531-3
Primary Topic
Syphilis Diagnosis and Treatment
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Diagnostic performance evaluation of the Abbott Bioline™ HIV/Syphilis Duo rapid test at the national public health laboratory, Yaounde-Cameroon

Philippe Salomon Nguwoh, Nafissatou Ibnou Moussa, Amira Moussa, Hamsatou Hadja et al.
BMC Infectious Diseases
Syphilis Diagnosis and Treatment
article

Diagnostic performance evaluation of the Abbott Bioline™ HIV/Syphilis Duo rapid test at the national public health laboratory, Yaounde-Cameroon

Philippe Salomon Nguwoh, Nafissatou Ibnou Moussa, Amira Moussa, Hamsatou Hadja, René Ghislain Essomba, Marie Claire Assoumou Okomo, Victor N. Fondoh, Amina Al-Mayé Bit Younouss, Désiré Tchoffo, Ibrahima Halilou, Emile K. Yuosembom, Youssoufa Taoufick, Blaise Akenji Mboringong, Marcel Mbarga Foe, Emmanuel Orock Eta, Guianni Mpiwouo Panyere, Jérôme Ngantchui Tchuisseu, John Francois Ndombol Tembten, Constance Nyanda Nyeng, Dodo Nyako Balkissou, Cressence Fouda, Solange Pefouakeu Pepa, Giraud Donfack Ngueguim, Christian Taheu Ngounouh, Lydie Nyatte, Henri Lucien Kamga
article en

Abstract

The Bioline™ HIV/Syphilis Duo rapid diagnostic test (RDT) is the first World Health Organization (WHO) prequalified Duo RDT for the simultaneous detection of human immunodeficiency virus (HIV) and Treponema pallidum antibodies in human blood samples. Several studies have shown the satisfactory diagnostic performance of this test among vulnerable populations such as pregnant women. However, the diagnostic performance of this Duo test is scarce in the general population in Cameroon. The present study aimed to assess the diagnostic performance of the Abbot Bioline™ HIV/Syphilis Duo RDT in Cameroon. A laboratory based cross-sectional and comparative study was conducted from January, 6th to June, 19th 2024 on serum/plasma collected from adult individuals (aged: ≥ 21 years), recruited consecutively in health facilities in eight regions of Cameroon. The blood samples were received and stored at the National Public Health Laboratory (NPHL). The Bioline™ HIV 1/2 test was performed following the national testing algorithm with Determine™ HIV 1/2 (first test) and KHB Shanghai HIV1/2 (second test). Meanwhile, the performance of the Bioline™ syphilis test was compared with the Hightop RDT and Treponema pallidum hemagglutination assay (TPHA). The enzyme linked immunosorbent assay (ELISA) was used as the reference assay. Sensitivity, specificity, positive predictive value (PPV), negative predictive value (NPV), accuracy and Cohen’s kappa (κ) value were calculated for each marker against the reference assay within a 95% confidence interval (CI) using Epi-info version 7, MedCalc and QuickCalcs for online statistical analysis. A total of 534 blood samples were tested at the NPHL with 70.97% (379/534) from females and 29.03% (155/534) from males. For HIV diagnosis, the Bioline™ HIV1/2 test had a sensitivity of 99.70% (95% CI: 98.30–99.9), specificity of 98.5% (95% CI: 95.58–99.5), PPV of 99.1% (95% CI: 97.40–99.7), NPV of 99.5% (95% CI: 97.3–99.9), accuracy of 99.1%(95% CI: 97-99.7) and Cohen’s kappa of 98.4%(95% CI:96.7–99.9). For syphilis diagnosis, the Bioline™ test had a sensitivity of 95% (95% CI: 73.7–99.7), specificity of 100% (95% CI: 99.1–100), PPV of 100% (95% CI: 79.1–100), NPV of 99.8% (95% CI: 98.7–100), accuracy of 99.9% (95% CI: 99–100), and Cohen’s kappa of 97.3%(95% CI:92.1–100). The results of this study show that the Abbot Bioline™ HIV/Syphilis Duo RDT has a good performance. Therefore, we suggest that this test could be suitable for use in the diagnosis of HIV and syphilis.

BMC Infectious DiseasesVol. 26(1)
Université de Yaoundé I (CM), University of Douala (CM), Université de Dschang (CM), University of Maroua (CM), University of Bamenda (CM), Limbe Provincial Hospital (CM), University of Kinshasa (CD), Bamenda University of Science and Technology (CM), Mali-Folkecenter (ML), Ministry of Higher Education (CM), Institut de Recherches Médicales et d’Etudes des Plantes Médicinales (CM), Ministry of Public Health (CM), Catholic University of Central Africa (CM), Andalas University (ID)
Good health and well-being
Openalex Percentile: Top 65%
Syphilis Diagnosis and Treatment
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.