A Breakthrough SIA-Based Dual Assay for Simultaneous Evaluation of Antioxidant Capacity via ABTS and FRAP Mechanisms

An innovative method, termed FRAP/ABTS-SIA, was developed to simultaneously integrate the FRAP and ABTS antioxidant assays within a single sequential injection analysis (SIA) system with spectrophotometric detection. Leveraging the kinetic differences between the assays and controlling the dispersion, a compact aspiration sequence (antioxidant-FRAP-ABTS-antioxidant-water) was optimized using a central composite design, defining a flow rate of 40 μL s<sup>-1</sup> and aspiration volumes of 43, 38, 38, 43, and 100 μL, respectively. The system incorporated a helical reaction coil positioned before the detector, allowing the antioxidant-FRAP bolus to react while the ABTS-antioxidant-water sequence was aspirated into the holding coil. This configuration enhanced the FRAP signal and enabled clear separation of both analytical responses. Compared to conventional batch protocols, this strategy reduced FRAP reagent concentrations by 70% and ABTS<sup>•+</sup> radical concentrations by 50%. The method delivers responses within a 2 min run, achieving a throughput of ∼30 samples h<sup>-1</sup>. Linearity was confirmed for both assays over the range 10-120 μmol L<sup>-1</sup> Trolox, with detection limits of 0.031 μmol L<sup>-1</sup> (FRAP) and 0.0047 μmol L<sup>-1</sup> (ABTS). Intralaboratory precision was below 2% RSD, and recoveries ranged from 97.3 to 106.2% (FRAP) and 92.8 to 105.4% (ABTS). The method was successfully applied to complex food matrices─including coffees, wines, juices, and spices─showing correlations ≥0.99 with microplate reference assays. High-throughput, reagent savings, metrological robustness, and simplified data processing position FRAP/ABTS-SIA as an efficient and reliable tool for routine antioxidant capacity evaluation in food and biomedical applications.

Authors

Institutions

Publication Details

Journal
Analytical Chemistry
Published
2026-02-04
DOI
https://doi.org/10.1021/acs.analchem.5c05489
Primary Topic
Electrochemical sensors and biosensors
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

A Breakthrough SIA-Based Dual Assay for Simultaneous Evaluation of Antioxidant Capacity via ABTS and FRAP Mechanisms

María del Pilar Cañizares-Macías, Willmann Antonio Jiménez Morales
Analytical Chemistry
Electrochemical sensors and biosensors
article

A Breakthrough SIA-Based Dual Assay for Simultaneous Evaluation of Antioxidant Capacity via ABTS and FRAP Mechanisms

María del Pilar Cañizares-Macías, Willmann Antonio Jiménez Morales
article en

Abstract

An innovative method, termed FRAP/ABTS-SIA, was developed to simultaneously integrate the FRAP and ABTS antioxidant assays within a single sequential injection analysis (SIA) system with spectrophotometric detection. Leveraging the kinetic differences between the assays and controlling the dispersion, a compact aspiration sequence (antioxidant-FRAP-ABTS-antioxidant-water) was optimized using a central composite design, defining a flow rate of 40 μL s<sup>-1</sup> and aspiration volumes of 43, 38, 38, 43, and 100 μL, respectively. The system incorporated a helical reaction coil positioned before the detector, allowing the antioxidant-FRAP bolus to react while the ABTS-antioxidant-water sequence was aspirated into the holding coil. This configuration enhanced the FRAP signal and enabled clear separation of both analytical responses. Compared to conventional batch protocols, this strategy reduced FRAP reagent concentrations by 70% and ABTS<sup>•+</sup> radical concentrations by 50%. The method delivers responses within a 2 min run, achieving a throughput of ∼30 samples h<sup>-1</sup>. Linearity was confirmed for both assays over the range 10-120 μmol L<sup>-1</sup> Trolox, with detection limits of 0.031 μmol L<sup>-1</sup> (FRAP) and 0.0047 μmol L<sup>-1</sup> (ABTS). Intralaboratory precision was below 2% RSD, and recoveries ranged from 97.3 to 106.2% (FRAP) and 92.8 to 105.4% (ABTS). The method was successfully applied to complex food matrices─including coffees, wines, juices, and spices─showing correlations ≥0.99 with microplate reference assays. High-throughput, reagent savings, metrological robustness, and simplified data processing position FRAP/ABTS-SIA as an efficient and reliable tool for routine antioxidant capacity evaluation in food and biomedical applications.

Analytical ChemistryVol. 98(6)
Universidad Nacional Autónoma de México (MX), Benito Juárez Autonomous University of Oaxaca (MX)
Openalex Percentile: Top 67%
Electrochemical sensors and biosensors
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.