Surface roughness and preload loss of implant-supported screw retained three-unit framework fabricated by different techniques

Aim This study evaluated the effects of selective laser melting (SLM), milling, and casting fabrication techniques on surface roughness and screw preload loss of three-unit screw-retained implant-supported Cobalt-Chromium (Co-Cr) frameworks. Methods Ten frame-works were fabricated by each fabrication method (n = 10), and the surface roughness (Ra) was measured using a 3D optical profilometer. The frameworks were torqued to 30 Ncm, retightened after 15 minutes, and subjected to 500,000 cycles in a dual-axis chewing simulator. Pre- and post-load reverse torque values (RTV) were recorded for molar and premolar, and the reverse torque difference (RTD) was calculated. Data were analyzed using ANOVA and Tukey-Kramer tests (α = 0.01). Correlation analyses were performed to determine any relation between surface roughness and preload (RTV and RTD) across all groups. Results Significant differences were found in surface roughness and preload loss among groups (p < 0.01). In molars, the milled group showed the lowest Ra (0.30 µm) and highest post-load RTV (25.2 Ncm), followed by the SLM group (1.73 µm, 20.2 Ncm) and cast group (1.65 µm, 16.9 Ncm). In premolars, the milled group again showed the lowest Ra (0.24 µm) and highest RTV (25.2 Ncm), followed by the SLM group (1.66 µm, 19.4 Ncm) and the cast group (2.57 µm, 17.4 Ncm). There was no significant correlation between roughness and preload loss. Conclusion Different manufacturing technique significantly affects surface topography and screw preload. Milled frameworks showed the least surface roughness and the most torque stability.

Authors

Institutions

Publication Details

Journal
Frontiers in Materials
Published
2026-02-04
DOI
https://doi.org/10.3389/fmats.2026.1763096
Primary Topic
Dental materials and restorations
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Surface roughness and preload loss of implant-supported screw retained three-unit framework fabricated by different techniques

Shahad Mohammmed Altuwaijri, Fahim Vohra, Refal Albaijan, Rawan Alsaif et al.
Frontiers in Materials
Dental materials and restorations
article

Surface roughness and preload loss of implant-supported screw retained three-unit framework fabricated by different techniques

Shahad Mohammmed Altuwaijri, Fahim Vohra, Refal Albaijan, Rawan Alsaif, Hussain D. Alsayed, Mohammed Alrabiah
article en

Abstract

Aim This study evaluated the effects of selective laser melting (SLM), milling, and casting fabrication techniques on surface roughness and screw preload loss of three-unit screw-retained implant-supported Cobalt-Chromium (Co-Cr) frameworks. Methods Ten frame-works were fabricated by each fabrication method (n = 10), and the surface roughness (Ra) was measured using a 3D optical profilometer. The frameworks were torqued to 30 Ncm, retightened after 15 minutes, and subjected to 500,000 cycles in a dual-axis chewing simulator. Pre- and post-load reverse torque values (RTV) were recorded for molar and premolar, and the reverse torque difference (RTD) was calculated. Data were analyzed using ANOVA and Tukey-Kramer tests (α = 0.01). Correlation analyses were performed to determine any relation between surface roughness and preload (RTV and RTD) across all groups. Results Significant differences were found in surface roughness and preload loss among groups (p < 0.01). In molars, the milled group showed the lowest Ra (0.30 µm) and highest post-load RTV (25.2 Ncm), followed by the SLM group (1.73 µm, 20.2 Ncm) and cast group (1.65 µm, 16.9 Ncm). In premolars, the milled group again showed the lowest Ra (0.24 µm) and highest RTV (25.2 Ncm), followed by the SLM group (1.66 µm, 19.4 Ncm) and the cast group (2.57 µm, 17.4 Ncm). There was no significant correlation between roughness and preload loss. Conclusion Different manufacturing technique significantly affects surface topography and screw preload. Milled frameworks showed the least surface roughness and the most torque stability.

Frontiers in MaterialsVol. 13
Nihon University (JP), Princess Nourah bint Abdulrahman University (SA), Prince Sattam Bin Abdulaziz University (SA), King Saud University (SA)
Openalex Percentile: Top 66%
Dental materials and restorations
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.