Extending Calibration Anchors Resolves Low- <i>m</i> / <i>z</i> Shifts in Negative-Mode Orbitrap Workflows

Reliable mass accuracy is essential for the confident identification of diagnostic fragment ions in high-resolution mass spectrometry. During the analysis of perfluoroalkyl sulfonic acids (PFSAs) with a Q Exactive Orbitrap, we consistently observed the characteristic O<sub>3</sub>Ṡ<sup>-</sup> fragment, expected at a mass-to-charge ratio (<i>m</i>/<i>z</i>) of 79.9574, reported at <i>m</i>/<i>z</i> 79.9568, a systematic deviation of approximately -6 parts per million (ppm). Similar errors affected other ions below <i>m</i>/<i>z</i> 100, while precursors and higher-<i>m</i>/<i>z</i> fragments remained within the specification. Comparison with an Exploris 120 instrument, calibrated with anchors down to <i>m</i>/<i>z</i> 59, confirmed that the deviation is not intrinsic to the ion but originates from the limited calibration range of the Q Exactive standard calibration solution. We demonstrate that extending calibration to include additional low-<i>m</i>/<i>z</i> anchors generated <i>in situ</i> by in-source fragmentation fully corrects the error without affecting the accuracy at a higher <i>m</i>/<i>z</i>. This adjustment resolves systematic deviations for ions below <i>m</i>/<i>z</i> 100 in the Q Exactive instruments.

Authors

Institutions

Publication Details

Journal
Journal of the American Society for Mass Spectrometry
Published
2026-02-03
DOI
https://doi.org/10.1021/jasms.5c00362
Primary Topic
Per- and polyfluoroalkyl substances research
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Extending Calibration Anchors Resolves Low- m / z Shifts in Negative-Mode Orbitrap Workflows

Francisco José Díaz-Galiano, Bruno Le Bizec
Journal of the American Society for Mass Spectrometry
Per- and polyfluoroalkyl substances research
article

Extending Calibration Anchors Resolves Low- m / z Shifts in Negative-Mode Orbitrap Workflows

Francisco José Díaz-Galiano, Bruno Le Bizec
article en

Abstract

Reliable mass accuracy is essential for the confident identification of diagnostic fragment ions in high-resolution mass spectrometry. During the analysis of perfluoroalkyl sulfonic acids (PFSAs) with a Q Exactive Orbitrap, we consistently observed the characteristic O<sub>3</sub>Ṡ<sup>-</sup> fragment, expected at a mass-to-charge ratio (<i>m</i>/<i>z</i>) of 79.9574, reported at <i>m</i>/<i>z</i> 79.9568, a systematic deviation of approximately -6 parts per million (ppm). Similar errors affected other ions below <i>m</i>/<i>z</i> 100, while precursors and higher-<i>m</i>/<i>z</i> fragments remained within the specification. Comparison with an Exploris 120 instrument, calibrated with anchors down to <i>m</i>/<i>z</i> 59, confirmed that the deviation is not intrinsic to the ion but originates from the limited calibration range of the Q Exactive standard calibration solution. We demonstrate that extending calibration to include additional low-<i>m</i>/<i>z</i> anchors generated <i>in situ</i> by in-source fragmentation fully corrects the error without affecting the accuracy at a higher <i>m</i>/<i>z</i>. This adjustment resolves systematic deviations for ions below <i>m</i>/<i>z</i> 100 in the Q Exactive instruments.

Journal of the American Society for Mass SpectrometryVol. 37(3)
Institut National de la Recherche Agronomique (MA), Oniris (FR)
Openalex Percentile: Top 71%
Per- and polyfluoroalkyl substances research
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.