Failure of 18F-florbetapir Amyloid PET Imaging after Injection With an Automated Injector Using an Air-vent Filter for FDG Injection

A 77-year-old woman with mild cognitive impairment was referred to 18F-florbetapir positron emission tomography (PET) to assess the indication of a disease-modifying drug against Alzheimer disease. PET images showed no uptake in the brain and only slight uptake in the sinus veins. In an automated injector, almost all the radioactivity was trapped in an air-vent filter, which is usually used for FDG injections and is made of mixed cellulose ester. After injection using other types of filters not made of mixed cellulose ester, 18F-florbetapir PET imaging was successful. The filter made of mixed cellulose ester should never be used for 18F-florbetapir injection.

Authors

Institutions

Publication Details

Journal
Clinical Nuclear Medicine
Published
2026-02-03
DOI
https://doi.org/10.1097/rlu.0000000000006273
Primary Topic
Intracerebral and Subarachnoid Hemorrhage Research
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Failure of 18F-florbetapir Amyloid PET Imaging after Injection With an Automated Injector Using an Air-vent Filter for FDG Injection

Yohji Matsusaka, Kaima Suzuki, Ichiei Kuji, Akira Seto et al.
Clinical Nuclear Medicine
Intracerebral and Subarachnoid Hemorrhage Research
article

Failure of 18F-florbetapir Amyloid PET Imaging after Injection With an Automated Injector Using an Air-vent Filter for FDG Injection

Yohji Matsusaka, Kaima Suzuki, Ichiei Kuji, Akira Seto, Yuji Kato
article en

Abstract

A 77-year-old woman with mild cognitive impairment was referred to 18F-florbetapir positron emission tomography (PET) to assess the indication of a disease-modifying drug against Alzheimer disease. PET images showed no uptake in the brain and only slight uptake in the sinus veins. In an automated injector, almost all the radioactivity was trapped in an air-vent filter, which is usually used for FDG injections and is made of mixed cellulose ester. After injection using other types of filters not made of mixed cellulose ester, 18F-florbetapir PET imaging was successful. The filter made of mixed cellulose ester should never be used for 18F-florbetapir injection.

Clinical Nuclear MedicineVol. 51(3)
Suzuki (Japan) (JP), Saitama Medical University (JP)
Good health and well-being
Openalex Percentile: Top 69%
Intracerebral and Subarachnoid Hemorrhage Research
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.