A novel projection data domain material decomposition method for dual-energy CT and its impact on the accuracy of attenuation values

<i>Objective.</i>Despite major advances in dual-energy computed tomography (CT), obtaining accurate attenuation values for quantitative applications remains a technical challenge. To address this topic, we introduce a novel projection data domain material decomposition method that is an extension of an approach we recently proposed for beam hardening correction in single energy CT.<i>Approach.</i>The proposed method employs object-specific scatter correction and an analytical energy response model. We compare its performance to image-based material decomposition on accuracy of attenuation values using the American College of Radiology (ACR) CT accreditation phantom, scanned with consecutive low and high energy axial scans in centered and off-centered positions. Accuracy is assessed across the five inserts, and the images are analyzed for beam hardening artifacts and noise. Additionally, we assess the usefulness of object-specific scatter correction, and we assess performance over conventional data domain material decomposition and for anthropomorphic abdomen phantom imaging.<i>Main results.</i>In the ACR phantom, the proposed method yielded a significant improvement in accuracy of the attenuation values, particularly at low energies (<70 keV), and an important reduction in beam hardening artifacts. While similarly high accuracy was achieved for water, quantitative error within the non-water inserts was lower and more uniform across the 30-140 keV range, especially in the more challenging off-centered positioning of the phantom. Noise showed expected parabolic behavior, but with minimum at lower keV, which may be clinically advantageous. Object-specific scatter correction was shown to prevent major artifacts. Advantages over conventional data-domain decomposition clearly appeared when only a standard phantom is available to calibrate the latter. Lastly, the proposed method was shown to perform well, without any changes, in the more complex scenario of abdominal phantom imaging.<i>Significance.</i>This work demonstrates that data-based material decomposition using an analytical energy response model with object-specific scatter correction offers a promising pathway to improve accuracy of CT attenuation values.

Authors

Institutions

Publication Details

Journal
Physics in Medicine and Biology
Published
2026-02-03
DOI
https://doi.org/10.1088/1361-6560/ae4163
Primary Topic
Advanced X-ray and CT Imaging
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

A novel projection data domain material decomposition method for dual-energy CT and its impact on the accuracy of attenuation values

F. Noo, Karl Stierstorfer, Viktor Haase, M McNitt‐Gray et al.
Physics in Medicine and Biology
Advanced X-ray and CT Imaging
article

A novel projection data domain material decomposition method for dual-energy CT and its impact on the accuracy of attenuation values

F. Noo, Karl Stierstorfer, Viktor Haase, M McNitt‐Gray, Andreas Maier
article en

Abstract

<i>Objective.</i>Despite major advances in dual-energy computed tomography (CT), obtaining accurate attenuation values for quantitative applications remains a technical challenge. To address this topic, we introduce a novel projection data domain material decomposition method that is an extension of an approach we recently proposed for beam hardening correction in single energy CT.<i>Approach.</i>The proposed method employs object-specific scatter correction and an analytical energy response model. We compare its performance to image-based material decomposition on accuracy of attenuation values using the American College of Radiology (ACR) CT accreditation phantom, scanned with consecutive low and high energy axial scans in centered and off-centered positions. Accuracy is assessed across the five inserts, and the images are analyzed for beam hardening artifacts and noise. Additionally, we assess the usefulness of object-specific scatter correction, and we assess performance over conventional data domain material decomposition and for anthropomorphic abdomen phantom imaging.<i>Main results.</i>In the ACR phantom, the proposed method yielded a significant improvement in accuracy of the attenuation values, particularly at low energies (<70 keV), and an important reduction in beam hardening artifacts. While similarly high accuracy was achieved for water, quantitative error within the non-water inserts was lower and more uniform across the 30-140 keV range, especially in the more challenging off-centered positioning of the phantom. Noise showed expected parabolic behavior, but with minimum at lower keV, which may be clinically advantageous. Object-specific scatter correction was shown to prevent major artifacts. Advantages over conventional data-domain decomposition clearly appeared when only a standard phantom is available to calibrate the latter. Lastly, the proposed method was shown to perform well, without any changes, in the more complex scenario of abdominal phantom imaging.<i>Significance.</i>This work demonstrates that data-based material decomposition using an analytical energy response model with object-specific scatter correction offers a promising pathway to improve accuracy of CT attenuation values.

Physics in Medicine and BiologyVol. 71(4)
University of California, Los Angeles (US), Friedrich-Alexander-Universität Erlangen-Nürnberg (DE), University of Utah (US), University of California System (US), Siemens Healthineers (Germany) (DE)
Affordable and clean energy
Openalex Percentile: Top 69%
Advanced X-ray and CT Imaging
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.