Interference Assessment for High‐Performance Liquid Chromatography HbA1c Assays

Interference is a significant source of laboratory errors that can cause erroneous measurement results. In clinical practice, all the measured results should exclude the influence of interference. In this study, the ion exchange high-performance liquid chromatography of hemoglobin A1c (HbA1c) was evaluated via three brands of HbA1c analyzers, and 12 endogenous and 24 exogenous substances were added to hemoglobin solutions with different HbA1c levels to simulate the corresponding clinical samples. After the potential interfering substances were screened via interference screening testing, the interference concentration thresholds of each potential interfering substance were determined on the basis of dose-response experiment. When these substances exceed the interference concentration thresholds, elevated HbA1c values are caused by endogenous cholic acid and low pH, whereas exogenous ethanol increases HbA1c measurements in samples with low values. However, HbA1c measurements decreased in samples with high values. Although there are some differences in the thresholds of interferent concentrations between different brands of analyzers, the differences are not significant. On the basis of the detailed interference concentration thresholds and interference substances, this evidence can be provided to clinical laboratory experts and help them interpret and determine the testing results of clinical laboratories.

Authors

Institutions

Publication Details

Journal
Biomedical Chromatography
Published
2026-02-03
DOI
https://doi.org/10.1002/bmc.70382
Primary Topic
Clinical Laboratory Practices and Quality Control
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Interference Assessment for High‐Performance Liquid Chromatography HbA1c Assays

Hongying Wang, Shengjin Cui, Jingang Luo, Nan Yan et al.
Biomedical Chromatography
Clinical Laboratory Practices and Quality Control
article

Interference Assessment for High‐Performance Liquid Chromatography HbA1c Assays

Hongying Wang, Shengjin Cui, Jingang Luo, Nan Yan, Yaxin Cao, Linghao Zhao, Wei Yang
article en

Abstract

Interference is a significant source of laboratory errors that can cause erroneous measurement results. In clinical practice, all the measured results should exclude the influence of interference. In this study, the ion exchange high-performance liquid chromatography of hemoglobin A1c (HbA1c) was evaluated via three brands of HbA1c analyzers, and 12 endogenous and 24 exogenous substances were added to hemoglobin solutions with different HbA1c levels to simulate the corresponding clinical samples. After the potential interfering substances were screened via interference screening testing, the interference concentration thresholds of each potential interfering substance were determined on the basis of dose-response experiment. When these substances exceed the interference concentration thresholds, elevated HbA1c values are caused by endogenous cholic acid and low pH, whereas exogenous ethanol increases HbA1c measurements in samples with low values. However, HbA1c measurements decreased in samples with high values. Although there are some differences in the thresholds of interferent concentrations between different brands of analyzers, the differences are not significant. On the basis of the detailed interference concentration thresholds and interference substances, this evidence can be provided to clinical laboratory experts and help them interpret and determine the testing results of clinical laboratories.

Biomedical ChromatographyVol. 40(3)
Xiamen University (CN), First Affiliated Hospital of Xiamen University (CN), Tianjin Special Equipment Supervision and Inspection Technology Research Institute (CN), University of Hong Kong - Shenzhen Hospital (CN), Tianjin Medical University (CN), University of Hong Kong (HK), University of South China (CN)
Openalex Percentile: Top 67%
Clinical Laboratory Practices and Quality Control
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.