Acute hemodynamic effects of a novel algorithm for cardiac resynchronization therapy optimization: Results from the BIO|Adapt study

<h2>ABSTRACT</h2><h3>Background</h3> A considerable proportion of heart failure patients with reduced left ventricular ejection fraction (LVEF) and prolonged QRS duration fail to respond adequately to Cardiac resynchronization therapy (CRT). Suboptimal atrioventricular (AV) delay programming is a key factor leading to diminished response. <h3>Objective</h3> To acutely correlate aortic velocity time integral (AoVTI), LVEF, and QRS duration with three programming strategies in the post-implantation setting: AV delay automatically set by a novel CRT AutoAdapt algorithm, AV delay optimized by echocardiography, and standard (‘in-box') programming. <h3>Methods</h3> Under normal intrinsic AV conduction (<250 ms after atrial pacing), the CRT AutoAdapt algorithm automatically switches to left ventricular-only pacing and adapts the AV delay to 50%, 70% (default), or 90% of intrinsic AV conduction. In our study, 163 CRT-defibrillator recipients (mean age 69.2±9.2 years, 32% female; LVEF 27.7%±6.6%; 76% with left bundle branch block) underwent echocardiographic and electrocardiographic assessments at pre-hospital discharge. <h3>Results</h3> With the default 70% AV adaptation, the CRT AutoAdapt automatically set similar AV delay (117.1±23.9 ms) compared to echocardiography-guided optimization (118.3±30.2 ms, p=0.94). Strong concordance was observed between the algorithm and echocardiographic optimization for AoVTI (concordance correlation coefficient: 0.96) and LVEF (0.93). Both optimization methods significantly outperformed standard AV programming for LVEF improvement (p<0.001) and QRS width reduction (p<0.01), with no significant differences between optimization approaches. <h3>Conclusion</h3> Acute hemodynamic effects of CRT AutoAdapt and echocardiographic AV delay optimization were similar, with equivalent cardiac performance. This device-based approach offers the potential advantage of continuous adaptation to changing physiological conditions without requiring resource-intensive in-office optimization procedures.

Authors

Institutions

Publication Details

Journal
Heart Rhythm O2
Published
2026-02-01
DOI
https://doi.org/10.1016/j.hroo.2026.01.030
Primary Topic
Cardiac pacing and defibrillation studies
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Acute hemodynamic effects of a novel algorithm for cardiac resynchronization therapy optimization: Results from the BIO|Adapt study

Jürgen Meyhöfer, Jun Kishihara, Mónica Pardo-Fresno, Iftikhar O Ebrahim et al.
Heart Rhythm O2
Cardiac pacing and defibrillation studies
article

Acute hemodynamic effects of a novel algorithm for cardiac resynchronization therapy optimization: Results from the BIO|Adapt study

Jürgen Meyhöfer, Jun Kishihara, Mónica Pardo-Fresno, Iftikhar O Ebrahim, Naoya Hironobe, Francisco Javier García-Fernández, Antonio Berruezo, Juan Gabriel Martínez, Atsuhiko Yagishita, Mariona Pinart, Kenji Ando, Hidemori Hayashi, Ritsushi Kato, Alan Bulava, Lars Anneken, Joaquin Osca-Asensi, Rafael Salguero-Bodes, Przemysław Mitkowski, Pedro Marques, Javier Jimenez, Masahide Harada
article en

Abstract

<h2>ABSTRACT</h2><h3>Background</h3> A considerable proportion of heart failure patients with reduced left ventricular ejection fraction (LVEF) and prolonged QRS duration fail to respond adequately to Cardiac resynchronization therapy (CRT). Suboptimal atrioventricular (AV) delay programming is a key factor leading to diminished response. <h3>Objective</h3> To acutely correlate aortic velocity time integral (AoVTI), LVEF, and QRS duration with three programming strategies in the post-implantation setting: AV delay automatically set by a novel CRT AutoAdapt algorithm, AV delay optimized by echocardiography, and standard (‘in-box') programming. <h3>Methods</h3> Under normal intrinsic AV conduction (<250 ms after atrial pacing), the CRT AutoAdapt algorithm automatically switches to left ventricular-only pacing and adapts the AV delay to 50%, 70% (default), or 90% of intrinsic AV conduction. In our study, 163 CRT-defibrillator recipients (mean age 69.2±9.2 years, 32% female; LVEF 27.7%±6.6%; 76% with left bundle branch block) underwent echocardiographic and electrocardiographic assessments at pre-hospital discharge. <h3>Results</h3> With the default 70% AV adaptation, the CRT AutoAdapt automatically set similar AV delay (117.1±23.9 ms) compared to echocardiography-guided optimization (118.3±30.2 ms, p=0.94). Strong concordance was observed between the algorithm and echocardiographic optimization for AoVTI (concordance correlation coefficient: 0.96) and LVEF (0.93). Both optimization methods significantly outperformed standard AV programming for LVEF improvement (p<0.001) and QRS width reduction (p<0.01), with no significant differences between optimization approaches. <h3>Conclusion</h3> Acute hemodynamic effects of CRT AutoAdapt and echocardiographic AV delay optimization were similar, with equivalent cardiac performance. This device-based approach offers the potential advantage of continuous adaptation to changing physiological conditions without requiring resource-intensive in-office optimization procedures.

Heart Rhythm O2
Poznan University of Medical Sciences (PL), Tokai University (JP), Hess (United States) (US), Friedrich-Alexander-Universität Erlangen-Nürnberg (DE), University Hospital Complex Of Vigo (ES), Steve Biko Hospital (ZA), Hospital Universitari i Politècnic La Fe (ES), University of South Bohemia in České Budějovice (CZ), Juntendo University (JP), Research Institute Hospital 12 de Octubre (ES), Hospital Quirón Teknon (ES), Centro de Investigación en Red en Enfermedades Cardiovasculares (ES), Hospital Universitario 12 De Octubre (ES), Hiroshima Prefectural Hospital (JP), Centro Hospitalar Lisboa Norte (PT), Hospital Universitario de Burgos (ES), Fujita Health University Hospital (JP), Nemocnice České Budějovice (CZ), Hospital General Universitario De Valencia (ES), Saitama International Medical Center (JP), Hospital General Universitario de Alicante Doctor Balmis (ES), Hospital de Santa Maria (PT), Biotronik (Germany) (DE), Kokura Memorial Hospital (JP), Instituto de Investigación Sanitaria La Fe (ES), Kitasato University (JP), Saitama Medical University (JP)
Openalex Percentile: Top 68%
Cardiac pacing and defibrillation studies
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.