Energy performance and environmental assessment of a commercial ULT freezer retrofitted with low-GWP refrigerants and electronic expansion valves

Operational challenges faced by ultra-low temperature (ULT) systems have led to their exclusion from environmental policies, resulting in insufficient market development. Consequently, currently used refrigerants have a high global warming potential (GWP), and technologies remain highly inefficient. Hence, this article experimentally analyses the operation of a commercial two-stage cascade ULT freezer retrofitted with A1 low-GWP refrigerants and electronic expansion valves. R448A is proposed for the high-temperature stage (HTS), and R469A, R472B, and R473A are proposed for the low-temperature stage (LTS) to replace the baseline refrigerants R404A (HTS) and R23 (LTS). Furthermore, the use of electronic expansion valves is compared to that of capillary tubes. The investigation initially focuses on the pull-down operation, in which the temperature varies from 20 °C down to −70 °C. After that, the hysteresis operation with freezer temperatures of −70 °C, −60 °C, and −50 °C is studied. The pull-down time of alternative refrigerants is higher than the baseline due to a reduced cooling power at the lowest temperatures. Regarding the hysteresis operation, the highest coefficient of performance (COP) is achieved by the R448A/R472B and R448A/R469A refrigerant pairs, which are up to 22.14% and 17.09% higher than the R404A/R23 pair, respectively. Moreover, R448A/R473A exhibits the highest cooling power, up to 73.18% higher than R404A/R23. Finally, the carbon footprint analysis reveals that all the proposed refrigerants significantly reduce (50% to 75%) the equivalent CO 2 emissions compared to the baseline system. Therefore, the solutions proposed in this work enable a reduction in environmental impact while increasing the energy efficiency of ULT systems.

Authors

Institutions

Publication Details

Journal
International Journal of Refrigeration
Published
2026-02-03
DOI
https://doi.org/10.1016/j.ijrefrig.2026.02.003
Primary Topic
Refrigeration and Air Conditioning Technologies
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Energy performance and environmental assessment of a commercial ULT freezer retrofitted with low-GWP refrigerants and electronic expansion valves

Adrián Mota-Babiloni, Joaquín Navarro-Esbrí, Cosmin-Mihai Udroiu, Ángel Barragán-Cervera et al.
International Journal of Refrigeration
Refrigeration and Air Conditioning Technologies
article

Energy performance and environmental assessment of a commercial ULT freezer retrofitted with low-GWP refrigerants and electronic expansion valves

Adrián Mota-Babiloni, Joaquín Navarro-Esbrí, Cosmin-Mihai Udroiu, Ángel Barragán-Cervera, Pau Giménez-Prades
article en

Abstract

Operational challenges faced by ultra-low temperature (ULT) systems have led to their exclusion from environmental policies, resulting in insufficient market development. Consequently, currently used refrigerants have a high global warming potential (GWP), and technologies remain highly inefficient. Hence, this article experimentally analyses the operation of a commercial two-stage cascade ULT freezer retrofitted with A1 low-GWP refrigerants and electronic expansion valves. R448A is proposed for the high-temperature stage (HTS), and R469A, R472B, and R473A are proposed for the low-temperature stage (LTS) to replace the baseline refrigerants R404A (HTS) and R23 (LTS). Furthermore, the use of electronic expansion valves is compared to that of capillary tubes. The investigation initially focuses on the pull-down operation, in which the temperature varies from 20 °C down to −70 °C. After that, the hysteresis operation with freezer temperatures of −70 °C, −60 °C, and −50 °C is studied. The pull-down time of alternative refrigerants is higher than the baseline due to a reduced cooling power at the lowest temperatures. Regarding the hysteresis operation, the highest coefficient of performance (COP) is achieved by the R448A/R472B and R448A/R469A refrigerant pairs, which are up to 22.14% and 17.09% higher than the R404A/R23 pair, respectively. Moreover, R448A/R473A exhibits the highest cooling power, up to 73.18% higher than R404A/R23. Finally, the carbon footprint analysis reveals that all the proposed refrigerants significantly reduce (50% to 75%) the equivalent CO 2 emissions compared to the baseline system. Therefore, the solutions proposed in this work enable a reduction in environmental impact while increasing the energy efficiency of ULT systems.

International Journal of RefrigerationVol. 185
Universitat Jaume I (ES)
Affordable and clean energy
Openalex Percentile: Top 69%
Refrigeration and Air Conditioning Technologies
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.