Paired assessment of two electromyographic neuromuscular monitors: stimpod NMX450X versus Datex-Ohmeda E-NMT

Quantitative neuromuscular monitoring (NMM) is widely recommended, yet its clinical adoption remains limited, partly due to usability issues. This study aimed to compare the clinical performance of two electromyography (EMG)-based NMM devices—the Datex-Ohmeda E-NMT module and the Stimpod NMS450X—under real-world operating room conditions. Twenty-four adult patients undergoing elective non-cardiac surgery were monitored simultaneously with both devices, applied contralaterally. Train-of-four (TOF) counts and TOF ratios were recorded at one-minute intervals. Agreement between devices was assessed using Bland–Altman analysis, repeatability coefficients, and the concordance correlation coefficient (CCC). The Datex-Ohmeda recorded an average of 121 TOF-count and 71 TOF-ratio values per patient, compared with 56 and 39, respectively, for the Stimpod. Inter-device correspondence was low, with numerous missing or physiologically implausible values on the Stimpod stream. The repeatability coefficient was 0.81 (95% CI: 0.80–0.83), indicating poor precision. The CCC was 0.662, consistent with weak agreement. In routine clinical use, the Stimpod NMS450X and Datex-Ohmeda E-NMT demonstrated poor concordance and high variability, likelyinfluenced by device-specific artifacts, connectivity limitations and electrode sensitivity.These findings highlight the need for device-specific validation of EMG NMM, both for clinical practice and for research purposes.

Authors

Institutions

Publication Details

Journal
Journal of Clinical Monitoring and Computing
Published
2026-02-03
DOI
https://doi.org/10.1007/s10877-026-01411-5
Primary Topic
Intraoperative Neuromonitoring and Anesthetic Effects
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Paired assessment of two electromyographic neuromuscular monitors: stimpod NMX450X versus Datex-Ohmeda E-NMT

Domien Vanhonacker, Hugo Carvalho, Wilfried Cools, Ulrich Janssens et al.
Journal of Clinical Monitoring and Computing
Intraoperative Neuromonitoring and Anesthetic Effects
article

Paired assessment of two electromyographic neuromuscular monitors: stimpod NMX450X versus Datex-Ohmeda E-NMT

Domien Vanhonacker, Hugo Carvalho, Wilfried Cools, Ulrich Janssens, Manish Zambre, Jarne De Mey
article en

Abstract

Quantitative neuromuscular monitoring (NMM) is widely recommended, yet its clinical adoption remains limited, partly due to usability issues. This study aimed to compare the clinical performance of two electromyography (EMG)-based NMM devices—the Datex-Ohmeda E-NMT module and the Stimpod NMS450X—under real-world operating room conditions. Twenty-four adult patients undergoing elective non-cardiac surgery were monitored simultaneously with both devices, applied contralaterally. Train-of-four (TOF) counts and TOF ratios were recorded at one-minute intervals. Agreement between devices was assessed using Bland–Altman analysis, repeatability coefficients, and the concordance correlation coefficient (CCC). The Datex-Ohmeda recorded an average of 121 TOF-count and 71 TOF-ratio values per patient, compared with 56 and 39, respectively, for the Stimpod. Inter-device correspondence was low, with numerous missing or physiologically implausible values on the Stimpod stream. The repeatability coefficient was 0.81 (95% CI: 0.80–0.83), indicating poor precision. The CCC was 0.662, consistent with weak agreement. In routine clinical use, the Stimpod NMS450X and Datex-Ohmeda E-NMT demonstrated poor concordance and high variability, likelyinfluenced by device-specific artifacts, connectivity limitations and electrode sensitivity.These findings highlight the need for device-specific validation of EMG NMM, both for clinical practice and for research purposes.

Journal of Clinical Monitoring and Computing
Vrije Universiteit Brussel (BE), Onze Lieve Vrouwziekenhuis Hospital (BE), AZ Sint-Jan (BE), Universitair Ziekenhuis Brussel (BE)
No poverty
Openalex Percentile: Top 69%
Intraoperative Neuromonitoring and Anesthetic Effects
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

Paired assessment of two electromyographic neuromuscular monitors: stimpod NMX450X versus Datex-Ohmeda E-NMT — Domien Vanhonacker, Hugo Carvalho, et al. · Journal of Clinical Monitoring and Computing (2026) | TGRS Research Map | TGRS