Establishment of a low dose rate 133Ba γ-ray calibration field for photon energy in the 300–400 keV range

A γ-ray calibration field using a sealed 133Ba source for photon energy in the 300-400 keV range was established at the Japan Atomic Energy Agency to evaluate radiation protection devices under low dose rate conditions. To suppress low-energy γ-ray components, a 2-mm lead filter was applied. The photon fluence spectrum was measured using a CdTe spectrometer and an unfolding method, yielding a spectrum with a mean energy of 321 keV. Reference air kerma rates were determined using ionization chambers calibrated in a 137Cs field, with energy correction factors applied. Spectrometry using a CeBr₃ scintillation spectrometer combined with the G(E) function method was employed to validate these corrections. Spectrum-averaged conversion coefficients were derived from the measured spectrum, indicating equivalence to the N-400 radiation quality. This field provided air kerma rates from 0.22 to 2.13 μGy h-1, suitable for testing devices used in environmental monitoring and nuclear emergency response.

Authors

Institutions

Publication Details

Journal
Radiation Protection Dosimetry
Published
2026-01-05
DOI
https://doi.org/10.1093/rpd/ncag002
Primary Topic
Radiation Shielding Materials Analysis
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Establishment of a low dose rate 133Ba γ-ray calibration field for photon energy in the 300–400 keV range

Munehiko Kowatari, F. Sato, Tomoya TSUJI, Yoshihiko Tanimura et al.
Radiation Protection Dosimetry
Radiation Shielding Materials Analysis
article

Establishment of a low dose rate 133Ba γ-ray calibration field for photon energy in the 300–400 keV range

Munehiko Kowatari, F. Sato, Tomoya TSUJI, Yoshihiko Tanimura, Hiroshi Yoshitomi
article en

Abstract

A γ-ray calibration field using a sealed 133Ba source for photon energy in the 300-400 keV range was established at the Japan Atomic Energy Agency to evaluate radiation protection devices under low dose rate conditions. To suppress low-energy γ-ray components, a 2-mm lead filter was applied. The photon fluence spectrum was measured using a CdTe spectrometer and an unfolding method, yielding a spectrum with a mean energy of 321 keV. Reference air kerma rates were determined using ionization chambers calibrated in a 137Cs field, with energy correction factors applied. Spectrometry using a CeBr₃ scintillation spectrometer combined with the G(E) function method was employed to validate these corrections. Spectrum-averaged conversion coefficients were derived from the measured spectrum, indicating equivalence to the N-400 radiation quality. This field provided air kerma rates from 0.22 to 2.13 μGy h-1, suitable for testing devices used in environmental monitoring and nuclear emergency response.

Radiation Protection Dosimetry
Japan Atomic Energy Agency (JP), National Institutes for Quantum Science and Technology (JP), The University of Osaka (JP)
Affordable and clean energy
Openalex Percentile: Top 70%
Radiation Shielding Materials Analysis
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.