A Benchtop Evaluation of Cervical Collar Design and Strap Tension
Abstract: This article presents a novel experimental setup for objectively evaluating cervical collar design and fit by monitoring strap tension and interface pressures across the device-skin interface. This study evaluated four commercially available cervical collars, achieving repeatable findings (ICC > 0.93) that were sensitive to differences between collars and across strap tensions. For each collar, there was a significant relationship (r > 0.4, p < 0.005) between strap tension and interface pressure values, with pressures exceeding 100 mmHg for high strap tensions. Differences in the distribution of pressures and the relationship between strap tension and interface pressures highlight the need for these objective measures of design and fit for cervical collars to evaluate new and existing designs. Keywords: cervical collars, pressure ulcers, interface pressure
Authors
- Davide Filingeri (ORCID: https://orcid.org/0000-0001-5652-395X)
- Laurence Russell
- Liudi Jiang
- Peter Worsley
Institutions
- University of Southampton (GB)
Publication Details
- Journal
- Medical Devices Evidence and Research
- Published
- 2026-02-01
- DOI
- https://doi.org/10.2147/mder.s541594
- Primary Topic
- Cervical and Thoracic Myelopathy
- Type
- article
- Field-Weighted Citation Impact
- 0.00
Funders
- Engineering and Physical Sciences Research Council