Assessment of Response Accuracy and Correction Factors of OSL Dosimeters in Diagnostic Radiology: Revealing Dependencies With Different Beam Qualities

This study establishes a robust and clinically applicable calibration protocol for optically stimulated luminescence dosimeters (OSLDs) in diagnostic radiology, with the aim of improving the accuracy of patient dose assessment. A total of 144 OSLDs were systematically irradiated under controlled conditions to assess their dosimetric response across a wide range of tube voltages (40-150 kVp) and square field sizes (10 × 10 cm² to 30 × 30 cm²). The dosimeters exhibited a sensitivity variation of ±6.6%, with an average background dose of 0.0185 mGy. The experimental data revealed a high dependence of OSLD response on photon energy, with dose values increasing by a factor of 11.5, from 0.1393 mGy at 40 kVp to 1.6072 mGy at 150 kVp for a constant field size of 10 × 10 cm². A pronounced non-linear dose escalation was observed in the mid-kVp range (70-100 kVp), where dose measurements increased by 72-90% as field size expanded. Energy and geometry-specific correction factors were derived, showing significant variation with field size, reaching maximum values of 9.81 for the 30 × 30 cm² field at 150 kVp and 7.43 for the 10 × 10 cm² field under the same conditions. Additionally, notable discrepancies were observed between experimentally derived effective beam energies and reference values reported by the International Atomic Energy Agency (IAEA), highlighting the need for localized calibration standards. These findings contribute to the standardization of OSLD calibration protocols in diagnostic radiology and support their implementation for accurate patient dose monitoring in clinical settings.

Authors

Institutions

Publication Details

Journal
Health Physics
Published
2026-02-02
DOI
https://doi.org/10.1097/hp.0000000000002045
Primary Topic
Advanced Radiotherapy Techniques
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Assessment of Response Accuracy and Correction Factors of OSL Dosimeters in Diagnostic Radiology: Revealing Dependencies With Different Beam Qualities

Mohammed Talbi, Abdellah Khallouqi, H. Sekkat, M. Haddioui et al.
Health Physics
Advanced Radiotherapy Techniques
article

Assessment of Response Accuracy and Correction Factors of OSL Dosimeters in Diagnostic Radiology: Revealing Dependencies With Different Beam Qualities

Mohammed Talbi, Abdellah Khallouqi, H. Sekkat, M. Haddioui, A. Halimi, O. El rhazouani
article en

Abstract

This study establishes a robust and clinically applicable calibration protocol for optically stimulated luminescence dosimeters (OSLDs) in diagnostic radiology, with the aim of improving the accuracy of patient dose assessment. A total of 144 OSLDs were systematically irradiated under controlled conditions to assess their dosimetric response across a wide range of tube voltages (40-150 kVp) and square field sizes (10 × 10 cm² to 30 × 30 cm²). The dosimeters exhibited a sensitivity variation of ±6.6%, with an average background dose of 0.0185 mGy. The experimental data revealed a high dependence of OSLD response on photon energy, with dose values increasing by a factor of 11.5, from 0.1393 mGy at 40 kVp to 1.6072 mGy at 150 kVp for a constant field size of 10 × 10 cm². A pronounced non-linear dose escalation was observed in the mid-kVp range (70-100 kVp), where dose measurements increased by 72-90% as field size expanded. Energy and geometry-specific correction factors were derived, showing significant variation with field size, reaching maximum values of 9.81 for the 30 × 30 cm² field at 150 kVp and 7.43 for the 10 × 10 cm² field under the same conditions. Additionally, notable discrepancies were observed between experimentally derived effective beam energies and reference values reported by the International Atomic Energy Agency (IAEA), highlighting the need for localized calibration standards. These findings contribute to the standardization of OSLD calibration protocols in diagnostic radiology and support their implementation for accurate patient dose monitoring in clinical settings.

Health Physics
Instituts Supérieurs des Professions Infirmières et Techniques de Santé (MA), Université Mohammed VI des Sciences et de la Santé (MA), Université Hassan 1er (MA), Hospital del Mar Research Institute (ES)
Affordable and clean energy
Openalex Percentile: Top 69%
Advanced Radiotherapy Techniques
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.