Point-of-Care Bilirubin Testing in Neonates: Comparative Performance of Blood Gas Analysis and Transcutaneous Bilirubinometry

<b>Background</b>: Neonatal jaundice is a common condition with potentially severe complications such as bilirubin-induced neurological dysfunction and kernicterus. While serum bilirubin (SBR) remains the standard laboratory measurement, point-of-care methods, such as transcutaneous bilirubinometry (TcB) and blood gas analysers (BGAs), offer rapid, less invasive alternatives. Direct comparisons of their diagnostic accuracy remain limited. <b>Objective</b>: The aim of this study was to assess and compare diagnostic accuracy and clinical utility of TcB and BGA against SBR in neonatal hyperbilirubinaemia screening. <b>Methods</b>: This retrospective study included neonates (n = 221) with concurrent SBR, BGA, and TcB measurements (n = 333). Assessment was via Passing-Bablok regression, Bland-Altman analysis, and Spearman correlation. Diagnostic performance was evaluated against jaundice thresholds in phototherapy charts (≥95th percentile threshold). Subgroup analyses considered phototherapy status, haemoglobin concentration, and Fitzpatrick skin type. <b>Results</b>: BGA showed stronger agreement with SBR (R<sup>2</sup> = 0.88) than TcB (R<sup>2</sup> = 0.43). BGA remained accurate regardless of phototherapy or haemoglobin levels. TcB accuracy declined post-phototherapy with reduced predictive value in darker-skinned neonates (Fitzpatrick III-VI) and increased false discovery rates. Both methods demonstrated low sensitivity (45.8%) but high specificity (>95%) and negative predictive value (~91%) for clinically significant hyperbilirubinaemia. BGA had a higher diagnostic odds ratio (47.5) than TcB (19.3). When individual patient sequential SBR and BGA measurements were compared for jaundice tracking (n = 175), there was high correlation, (r = 0.971) with no statistical differences, and 50% of measurements achieving agreement within 10 μmol/L. <b>Conclusions</b>: BGA is a more reliable alternative to SBR than TcB, particularly in time-critical or resource-limited settings. While TcB remains a non-invasive screening tool, limited accuracy post-phototherapy and with darker skinned neonates indicate confirmatory SBR testing. These findings support the selective and context-aware use of BGA and TcB to optimise neonatal hyperbilirubinaemia management and reduce interventions.

Authors

Institutions

Publication Details

Journal
Healthcare
Published
2026-02-01
DOI
https://doi.org/10.3390/healthcare14030370
Primary Topic
Neonatal Health and Biochemistry
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Point-of-Care Bilirubin Testing in Neonates: Comparative Performance of Blood Gas Analysis and Transcutaneous Bilirubinometry

Wei Fan, Bincy Francis, David Tran, Kay Weng Choy et al.
Healthcare
Neonatal Health and Biochemistry
article

Point-of-Care Bilirubin Testing in Neonates: Comparative Performance of Blood Gas Analysis and Transcutaneous Bilirubinometry

Wei Fan, Bincy Francis, David Tran, Kay Weng Choy, Amy Surkitt, Rami Subhi, George Francis Dargaville, Andrew Xu
article en

Abstract

<b>Background</b>: Neonatal jaundice is a common condition with potentially severe complications such as bilirubin-induced neurological dysfunction and kernicterus. While serum bilirubin (SBR) remains the standard laboratory measurement, point-of-care methods, such as transcutaneous bilirubinometry (TcB) and blood gas analysers (BGAs), offer rapid, less invasive alternatives. Direct comparisons of their diagnostic accuracy remain limited. <b>Objective</b>: The aim of this study was to assess and compare diagnostic accuracy and clinical utility of TcB and BGA against SBR in neonatal hyperbilirubinaemia screening. <b>Methods</b>: This retrospective study included neonates (n = 221) with concurrent SBR, BGA, and TcB measurements (n = 333). Assessment was via Passing-Bablok regression, Bland-Altman analysis, and Spearman correlation. Diagnostic performance was evaluated against jaundice thresholds in phototherapy charts (≥95th percentile threshold). Subgroup analyses considered phototherapy status, haemoglobin concentration, and Fitzpatrick skin type. <b>Results</b>: BGA showed stronger agreement with SBR (R<sup>2</sup> = 0.88) than TcB (R<sup>2</sup> = 0.43). BGA remained accurate regardless of phototherapy or haemoglobin levels. TcB accuracy declined post-phototherapy with reduced predictive value in darker-skinned neonates (Fitzpatrick III-VI) and increased false discovery rates. Both methods demonstrated low sensitivity (45.8%) but high specificity (>95%) and negative predictive value (~91%) for clinically significant hyperbilirubinaemia. BGA had a higher diagnostic odds ratio (47.5) than TcB (19.3). When individual patient sequential SBR and BGA measurements were compared for jaundice tracking (n = 175), there was high correlation, (r = 0.971) with no statistical differences, and 50% of measurements achieving agreement within 10 μmol/L. <b>Conclusions</b>: BGA is a more reliable alternative to SBR than TcB, particularly in time-critical or resource-limited settings. While TcB remains a non-invasive screening tool, limited accuracy post-phototherapy and with darker skinned neonates indicate confirmatory SBR testing. These findings support the selective and context-aware use of BGA and TcB to optimise neonatal hyperbilirubinaemia management and reduce interventions.

HealthcareVol. 14(3)
The University of Melbourne (AU), Northern Health (AU), Melbourne Health (AU)
Good health and well-being
Openalex Percentile: Top 69%
Neonatal Health and Biochemistry
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.