Evaluation of agreement between point-of-care and blood gas analyzers for lactate measurement in critically ill dogs

This study aimed to evaluate the analytical agreement between a handheld point-of-care (POC) lactate meter (StatStrip Xpress®2 Lac/Hb/Hct, Nova Biomedical) and a standard blood gas analyzer (Stat Profile Prime Plus® VET, Nova Biomedical) for measuring plasma lactate concentrations in critically ill dogs. Fifty-seven client-owned dogs were admitted to the Emergency Unit and Critical Care Unit (CCU) of the Small Animal Hospital, Chulalongkorn University, enrolled. Blood samples were analyzed using both a handheld POC meter and a blood gas analyzer. To evaluate analytical agreement across a wide concentration range, data were categorized into low (1–4 mmol/L; n = 17), moderate (4–10 mmol/L; n = 28), and high (10–20 mmol/L; n = 12) lactate groups. Spearman’s correlation, Passing-Bablok regression, and Bland-Altman analysis were performed to assess the agreement between the two devices. One-way ANOVA was used to analyze differences among the three groups. A strong positive correlation was observed between the two devices (r = 0.99). Passing–Bablok regression demonstrated acceptable agreement, with slopes of 1.02, 0.94, 1.01, and 1.38 for the overall dataset, low, moderate, and high lactate groups, respectively. Bland–Altman analysis showed mean biases of 0.07, 0.02, 0.26, and –0.57 mmol/L for the overall dataset, low-, moderate-, and high-lactate groups, respectively, indicating that the mean differences between the instruments were within acceptable analytical limits across most lactate ranges. In contrast, one-way ANOVA revealed a significant difference between the highlactate group and the others. In conclusion, the handheld POC lactate meter demonstrated a high degree of agreement with the standard blood gas analyzer, particularly in the low- and moderate-lactate groups.

Authors

Publication Details

Journal
The Thai Journal of Veterinary Medicine
Published
2026-01-28
DOI
https://doi.org/10.56808/2985-1130.3916
Primary Topic
Clinical Laboratory Practices and Quality Control
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Evaluation of agreement between point-of-care and blood gas analyzers for lactate measurement in critically ill dogs

Vachira Hunprasit, Manisin Vanichwatanaramlouk, Wipa Sripodok, Chutirat Torsahakul et al.
The Thai Journal of Veterinary Medicine
Clinical Laboratory Practices and Quality Control
article

Evaluation of agreement between point-of-care and blood gas analyzers for lactate measurement in critically ill dogs

Vachira Hunprasit, Manisin Vanichwatanaramlouk, Wipa Sripodok, Chutirat Torsahakul, Pemika Bunrueang, Parnkaew Wannamala, Kittiya Simphaisarn
article en

Abstract

This study aimed to evaluate the analytical agreement between a handheld point-of-care (POC) lactate meter (StatStrip Xpress®2 Lac/Hb/Hct, Nova Biomedical) and a standard blood gas analyzer (Stat Profile Prime Plus® VET, Nova Biomedical) for measuring plasma lactate concentrations in critically ill dogs. Fifty-seven client-owned dogs were admitted to the Emergency Unit and Critical Care Unit (CCU) of the Small Animal Hospital, Chulalongkorn University, enrolled. Blood samples were analyzed using both a handheld POC meter and a blood gas analyzer. To evaluate analytical agreement across a wide concentration range, data were categorized into low (1–4 mmol/L; n = 17), moderate (4–10 mmol/L; n = 28), and high (10–20 mmol/L; n = 12) lactate groups. Spearman’s correlation, Passing-Bablok regression, and Bland-Altman analysis were performed to assess the agreement between the two devices. One-way ANOVA was used to analyze differences among the three groups. A strong positive correlation was observed between the two devices (r = 0.99). Passing–Bablok regression demonstrated acceptable agreement, with slopes of 1.02, 0.94, 1.01, and 1.38 for the overall dataset, low, moderate, and high lactate groups, respectively. Bland–Altman analysis showed mean biases of 0.07, 0.02, 0.26, and –0.57 mmol/L for the overall dataset, low-, moderate-, and high-lactate groups, respectively, indicating that the mean differences between the instruments were within acceptable analytical limits across most lactate ranges. In contrast, one-way ANOVA revealed a significant difference between the highlactate group and the others. In conclusion, the handheld POC lactate meter demonstrated a high degree of agreement with the standard blood gas analyzer, particularly in the low- and moderate-lactate groups.

The Thai Journal of Veterinary MedicineVol. 56(1)
Good health and well-being
Openalex Percentile: Top 70%
Clinical Laboratory Practices and Quality Control
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

Evaluation of agreement between point-of-care and blood gas analyzers for lactate measurement in critically ill dogs — Vachira Hunprasit, Manisin Vanichwatanaramlouk, et al. · The Thai Journal of Veterinary Medicine (2026) | TGRS Research Map | TGRS