ACCURACY OF X-RAY EXPOSURE PARAMETERS IN SOME RADIO-DIAGNOSTIC CENTRES IN KADUNA STATE, NIGERIA

Background: Dose levels in radio-diagnostic settings depend largely on equipment status and examination techniques. It is essential to carry out quality control (QC) on the x-ray systems to ensure proper optimization and the maintenance of quality for better diagnostic outcome, and to avoid the risk of injury. Objective: To carryout QC for the assessment of the accuracy of exposure parameters for the purpose of optimization of radiation protection. Methodology: Quality control measurements including tube output, kVp, mA and timer accuracy, linearity and reproducibility were carried out using a multi-function quality control kit, RMI, 181B and measuring tape. Results: The results of these measurements indicate that the tube output of the machines within the kVp range of 60 to 120 falls between 13.52 x 10-3 and 161.26 x 10-3 with a linear relationship between the kVp and tube output in all the hospitals. It also showed exposure parameters’ operation to be within tolerance limits except in a single hospital with aspects of out-of-range performance. Conclusion: All equipment under study functioned within safe limits but further investigation and corrective action is required in one of the centres.

Authors

Institutions

Publication Details

Journal
JRRS
Published
2025-12-29
DOI
https://doi.org/10.82547/jrrs.vol32no1.65
Primary Topic
Radiation Dose and Imaging
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

ACCURACY OF X-RAY EXPOSURE PARAMETERS IN SOME RADIO-DIAGNOSTIC CENTRES IN KADUNA STATE, NIGERIA

Abeab Christopher Ishiekwen, Abubakar Sadiq Aliyu
JRRS
Radiation Dose and Imaging
article

ACCURACY OF X-RAY EXPOSURE PARAMETERS IN SOME RADIO-DIAGNOSTIC CENTRES IN KADUNA STATE, NIGERIA

Abeab Christopher Ishiekwen, Abubakar Sadiq Aliyu
article en

Abstract

Background: Dose levels in radio-diagnostic settings depend largely on equipment status and examination techniques. It is essential to carry out quality control (QC) on the x-ray systems to ensure proper optimization and the maintenance of quality for better diagnostic outcome, and to avoid the risk of injury. Objective: To carryout QC for the assessment of the accuracy of exposure parameters for the purpose of optimization of radiation protection. Methodology: Quality control measurements including tube output, kVp, mA and timer accuracy, linearity and reproducibility were carried out using a multi-function quality control kit, RMI, 181B and measuring tape. Results: The results of these measurements indicate that the tube output of the machines within the kVp range of 60 to 120 falls between 13.52 x 10-3 and 161.26 x 10-3 with a linear relationship between the kVp and tube output in all the hospitals. It also showed exposure parameters’ operation to be within tolerance limits except in a single hospital with aspects of out-of-range performance. Conclusion: All equipment under study functioned within safe limits but further investigation and corrective action is required in one of the centres.

JRRS
Federal University Lafia (NG), Ahmadu Bello University Teaching Hospital (NG)
Good health and well-being
Openalex Percentile: Top 28%
Radiation Dose and Imaging
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.