ACCURACY OF X-RAY EXPOSURE PARAMETERS IN SOME RADIO-DIAGNOSTIC CENTRES IN KADUNA STATE, NIGERIA
Background: Dose levels in radio-diagnostic settings depend largely on equipment status and examination techniques. It is essential to carry out quality control (QC) on the x-ray systems to ensure proper optimization and the maintenance of quality for better diagnostic outcome, and to avoid the risk of injury. Objective: To carryout QC for the assessment of the accuracy of exposure parameters for the purpose of optimization of radiation protection. Methodology: Quality control measurements including tube output, kVp, mA and timer accuracy, linearity and reproducibility were carried out using a multi-function quality control kit, RMI, 181B and measuring tape. Results: The results of these measurements indicate that the tube output of the machines within the kVp range of 60 to 120 falls between 13.52 x 10-3 and 161.26 x 10-3 with a linear relationship between the kVp and tube output in all the hospitals. It also showed exposure parameters’ operation to be within tolerance limits except in a single hospital with aspects of out-of-range performance. Conclusion: All equipment under study functioned within safe limits but further investigation and corrective action is required in one of the centres.
Authors
- Abeab Christopher Ishiekwen
- Abubakar Sadiq Aliyu
Institutions
- Federal University Lafia (NG)
- Ahmadu Bello University Teaching Hospital (NG)
Publication Details
- Journal
- JRRS
- Published
- 2025-12-29
- DOI
- https://doi.org/10.82547/jrrs.vol32no1.65
- Primary Topic
- Radiation Dose and Imaging
- Type
- article
- Field-Weighted Citation Impact
- 0.00