Development and preliminary evaluation of a transducer guiding system for hand ultrasound: With a focus on reproducibility and operational stability

Various strategies have been proposed to reduce operator-dependent variability in musculoskeletal ultrasound, including mechanical stabilization techniques. However, their effects on image reproducibility and transducer handling remain unclear in anatomically complex regions such as hand. This two-phase study aimed to (1) develop and validate a transducer guiding system for hand ultrasound and (2) evaluate its influence on image reproducibility and operational stability through image-based analyses. In Phase I, test-retest reliability of the guiding system was examined by repeatedly measuring soft-tissue thickness in 30 healthy participants. Intraclass Correlation Coefficients (ICC) ranged from 0.766 to 0.948, demonstrating good to excellent reliability. In Phase II, 16 ultrasound users acquired images under handheld and guiding-system-assisted conditions based on predefined reference images. Image reproducibility was evaluated using Normalized Cross-Correlation (NCC) and ICC, while operational stability was assessed via cosine similarity derived from M-mode segments during the pre-capture period, with group comparisons performed using the Wilcoxon signed-rank test. Novice users showed substantial ICC improvement when using the guiding system (0.487-0.681), approaching the consistency observed in experienced users. Both novice and experienced groups displayed higher NCC and cosine similarity values with the guiding system, indicating improved reproducibility and operational stability. Overall, the guiding system enhanced image outcomes by providing standardized positioning and mechanical stabilization. These findings highlight its potential to improve consistency in hand ultrasound assessments and serve as a supportive tool for novice ultrasound training. The results further demonstrate the feasibility of mechanical stabilization in reducing operator-related variability and enhancing image consistency in hand ultrasound examinations.

Authors

Institutions

Publication Details

Journal
Proceedings of the Institution of Mechanical Engineers Part H Journal of Engineering in Medicine
Published
2026-01-30
DOI
https://doi.org/10.1177/09544119251414217
Primary Topic
Ultrasound Imaging and Elastography
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Development and preliminary evaluation of a transducer guiding system for hand ultrasound: With a focus on reproducibility and operational stability

Mu-Ting Wang, Li-Chieh Kuo, Yu-Chen Lin, Tian Yang et al.
Proceedings of the Institution of Mechanical Engineers Part H Journal of Engineering in Medicine
Ultrasound Imaging and Elastography
article

Development and preliminary evaluation of a transducer guiding system for hand ultrasound: With a focus on reproducibility and operational stability

Mu-Ting Wang, Li-Chieh Kuo, Yu-Chen Lin, Tian Yang, Hong-Wen Wu, Hsiu-Yun Hsu
article en

Abstract

Various strategies have been proposed to reduce operator-dependent variability in musculoskeletal ultrasound, including mechanical stabilization techniques. However, their effects on image reproducibility and transducer handling remain unclear in anatomically complex regions such as hand. This two-phase study aimed to (1) develop and validate a transducer guiding system for hand ultrasound and (2) evaluate its influence on image reproducibility and operational stability through image-based analyses. In Phase I, test-retest reliability of the guiding system was examined by repeatedly measuring soft-tissue thickness in 30 healthy participants. Intraclass Correlation Coefficients (ICC) ranged from 0.766 to 0.948, demonstrating good to excellent reliability. In Phase II, 16 ultrasound users acquired images under handheld and guiding-system-assisted conditions based on predefined reference images. Image reproducibility was evaluated using Normalized Cross-Correlation (NCC) and ICC, while operational stability was assessed via cosine similarity derived from M-mode segments during the pre-capture period, with group comparisons performed using the Wilcoxon signed-rank test. Novice users showed substantial ICC improvement when using the guiding system (0.487-0.681), approaching the consistency observed in experienced users. Both novice and experienced groups displayed higher NCC and cosine similarity values with the guiding system, indicating improved reproducibility and operational stability. Overall, the guiding system enhanced image outcomes by providing standardized positioning and mechanical stabilization. These findings highlight its potential to improve consistency in hand ultrasound assessments and serve as a supportive tool for novice ultrasound training. The results further demonstrate the feasibility of mechanical stabilization in reducing operator-related variability and enhancing image consistency in hand ultrasound examinations.

Proceedings of the Institution of Mechanical Engineers Part H Journal of Engineering in MedicineVol. 240(2)
Dayeh University (TW), National Cheng Kung University Hospital (TW), National Cheng Kung University (TW), Southern Taiwan University of Science and Technology (TW)
Openalex Percentile: Top 72%
Ultrasound Imaging and Elastography
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.