Numerical analysis of low-frequency electromagnetic field effects from three-phase transformer on coronary stents and cardiac tissues

The widespread clinical adoption of novel magnesium alloy coronary stents, combined with increasing densification of urban power transmission infrastructure, highlighted a significant research gap regarding the effects of power-frequency electromagnetic fields (EMFs) on these implants. This study employed field-circuit coupling numerical methods to simulate electromagnetic field exposure in simulated patients with implanted coronary stent positioned at various locations near a 200kVA three-phase transformer. The analysis focused on the distribution patterns of induced electromagnetic fields within both cardiac tissues and the stent, as well as the resultant Ampere forces acting on the stent. The results showed that the simulated patient directly beneath the three-phase transformer was exposed to the maximum electromagnetic radiation, but the magnetic flux density (Bmax) and the induced electric field intensity (Emax) of the cardiac tissue were lower than the public exposure limits of the International Commission on Non-Ionizing Radiation Protection (ICNIRP). The Bmax and Emax of the stent at the same position were 1.245 μT and 5.086 × 10-4 mV/m, respectively. The maximum Ampere force density of the stent in the Y- axis (perpendicular to the coronal plane) was 3.714 × 10-6 N/m3. The above findings indicate that, under the conditions of this simulation, a 200 kVA power transformer exerts minimal interference on the magnesium alloy stent and cardiac tissues. The magnetic flux density and induced electric field in the heart tissues, as well as the Ampere force acting on the magnesium alloy stent, all remain within established safety limits.

Authors

Institutions

Publication Details

Journal
PLoS ONE
Published
2026-01-30
DOI
https://doi.org/10.1371/journal.pone.0340031
Primary Topic
Electromagnetic Fields and Biological Effects
Type
article
Field-Weighted Citation Impact
0.00

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Numerical analysis of low-frequency electromagnetic field effects from three-phase transformer on coronary stents and cardiac tissues

Mai Lu, Jia-Yi Luo, Rui Tian, Jing-Jing Cui
PLoS ONE
Electromagnetic Fields and Biological Effects
article

Numerical analysis of low-frequency electromagnetic field effects from three-phase transformer on coronary stents and cardiac tissues

Mai Lu, Jia-Yi Luo, Rui Tian, Jing-Jing Cui
article en

Abstract

The widespread clinical adoption of novel magnesium alloy coronary stents, combined with increasing densification of urban power transmission infrastructure, highlighted a significant research gap regarding the effects of power-frequency electromagnetic fields (EMFs) on these implants. This study employed field-circuit coupling numerical methods to simulate electromagnetic field exposure in simulated patients with implanted coronary stent positioned at various locations near a 200kVA three-phase transformer. The analysis focused on the distribution patterns of induced electromagnetic fields within both cardiac tissues and the stent, as well as the resultant Ampere forces acting on the stent. The results showed that the simulated patient directly beneath the three-phase transformer was exposed to the maximum electromagnetic radiation, but the magnetic flux density (Bmax) and the induced electric field intensity (Emax) of the cardiac tissue were lower than the public exposure limits of the International Commission on Non-Ionizing Radiation Protection (ICNIRP). The Bmax and Emax of the stent at the same position were 1.245 μT and 5.086 × 10-4 mV/m, respectively. The maximum Ampere force density of the stent in the Y- axis (perpendicular to the coronal plane) was 3.714 × 10-6 N/m3. The above findings indicate that, under the conditions of this simulation, a 200 kVA power transformer exerts minimal interference on the magnesium alloy stent and cardiac tissues. The magnetic flux density and induced electric field in the heart tissues, as well as the Ampere force acting on the magnesium alloy stent, all remain within established safety limits.

PLoS ONEVol. 21(1)
Lanzhou Jiaotong University (CN), China Railway Fifth Survey and Design Institute Group (CN), China Railway First Survey and Design Institute Group Co. Ltd.
National Natural Science Foundation of China
Sustainable cities and communities
Openalex Percentile: Top 75%
Electromagnetic Fields and Biological Effects
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.