Time Course of Speech Perception Performance Related to Advanced Bionics V1 Device Revision
The 7-year survival rate of initial V1 devices is around 30% to 33%, with the anticipation of an even lower survival rate over a longer period. Although speech perception recovers over time, it may take as long as 9 months post-revision for patients to return to pre-failure levels. We recommend rigorous monitoring using speech perception testing, both during V1 integrity monitoring and post-revision, to understand the impact of device failure in individual patients, particularly for children.
Authors
- Elizabeth A. Kelly (ORCID: https://orcid.org/0000-0001-7335-6698)
- Jeffrey L. Simmons
- Eric R. Rodriguez
- Z. Ellen Peng
Institutions
- Purdue University West Lafayette (US)
- Boys Town National Research Hospital (US)
Publication Details
- Journal
- Otology & Neurotology
- Published
- 2026-01-30
- DOI
- https://doi.org/10.1097/mao.0000000000004847
- Primary Topic
- Ear Surgery and Otitis Media
- Type
- article
- Field-Weighted Citation Impact