An analytical framework for quantifying recoil and foreshortening in ultra-thin cobalt–chromium coronary stents

• Develops a coupled analytical-FEA framework for balloon-expandable thin-walled stents. • Derives closed-form expressions linking stent unit-cell geometry to axial deformation. • Incorporates nonlinear plasticity to accurately predict post-expansion stent behavior. • Validates optimized design achieving 2.7% foreshortening and controlled recoil within 7%. • Extends applicability of thin-walled lattice design to elastic-plastic biomedical structures. Ultrathin cobalt-chromium (Co-Cr) stents offer superior clinical outcomes but face inherent mechanical trade-offs, particularly regarding radial stiffness and recoil. While finite element analysis (FEA) is the standard for optimizing these designs, it is computationally expensive, and existing analytical models often fail to capture the complex elastoplastic deformation of balloon-expandable architectures. To address this gap, this study introduces a hinge-resolved, closed-form analytical framework for predicting the elastoplastic deployment behavior of balloon-expandable stents. The model explicitly links unit-cell geometry, including hinge radius, link length, and cell number, to critical performance metrics such as foreshortening and recoil, capturing both crimped and expanded configurations. Analytical predictions showed close agreement with nonlinear FEA across all configurations. Through parametric optimization, foreshortening was reduced from 17% in the baseline design to 2.7%, while recoil was maintained at 3.7%, well within FDA regulatory thresholds. These results demonstrate that increasing the circumferential cell count, link length and hinge radius minimizes foreshortening, whereas decreasing them reduces recoil. Overall, the proposed framework provides a rapid, interpretable, and computationally efficient tool for early-stage design optimization of laser-cut stents, effectively bridging the gap between analytical theory and complex numerical simulation.

Authors

Institutions

Publication Details

Journal
Results in Engineering
Published
2026-01-30
DOI
https://doi.org/10.1016/j.rineng.2026.109363
Primary Topic
Coronary Interventions and Diagnostics
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

An analytical framework for quantifying recoil and foreshortening in ultra-thin cobalt–chromium coronary stents

Youngjae Chun, Chang Hun Kum, Jae Cho, Moataz Elsisy et al.
Results in Engineering
Coronary Interventions and Diagnostics
article

An analytical framework for quantifying recoil and foreshortening in ultra-thin cobalt–chromium coronary stents

Youngjae Chun, Chang Hun Kum, Jae Cho, Moataz Elsisy, Ch.A.R. Saleh, Amr F. Mohamed, Gyuhyun Jin, Sang Hyun An
article en

Abstract

• Develops a coupled analytical-FEA framework for balloon-expandable thin-walled stents. • Derives closed-form expressions linking stent unit-cell geometry to axial deformation. • Incorporates nonlinear plasticity to accurately predict post-expansion stent behavior. • Validates optimized design achieving 2.7% foreshortening and controlled recoil within 7%. • Extends applicability of thin-walled lattice design to elastic-plastic biomedical structures. Ultrathin cobalt-chromium (Co-Cr) stents offer superior clinical outcomes but face inherent mechanical trade-offs, particularly regarding radial stiffness and recoil. While finite element analysis (FEA) is the standard for optimizing these designs, it is computationally expensive, and existing analytical models often fail to capture the complex elastoplastic deformation of balloon-expandable architectures. To address this gap, this study introduces a hinge-resolved, closed-form analytical framework for predicting the elastoplastic deployment behavior of balloon-expandable stents. The model explicitly links unit-cell geometry, including hinge radius, link length, and cell number, to critical performance metrics such as foreshortening and recoil, capturing both crimped and expanded configurations. Analytical predictions showed close agreement with nonlinear FEA across all configurations. Through parametric optimization, foreshortening was reduced from 17% in the baseline design to 2.7%, while recoil was maintained at 3.7%, well within FDA regulatory thresholds. These results demonstrate that increasing the circumferential cell count, link length and hinge radius minimizes foreshortening, whereas decreasing them reduces recoil. Overall, the proposed framework provides a rapid, interpretable, and computationally efficient tool for early-stage design optimization of laser-cut stents, effectively bridging the gap between analytical theory and complex numerical simulation.

Results in EngineeringVol. 29
Cairo University (EG), University of Pittsburgh (US), Osstem Implant (South Korea) (KR), University of Pittsburgh Medical Center (US), McGowan Institute for Regenerative Medicine (US), Daegu-Gyeongbuk Medical Innovation Foundation (KR)
Openalex Percentile: Top 72%
Coronary Interventions and Diagnostics
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines